Difference between revisions of "LFI design, qualification, and performance"

From Planck Legacy Archive Wiki
Jump to: navigation, search
(Systematic Effects)
 
(362 intermediate revisions by 11 users not shown)
Line 1: Line 1:
[[Category:Instruments]]
+
==<span id="LFIDescription">Instrument description</span>==
== Overview ==
+
The Low-Frequency Instrument (see Fig. 1 in the [[LFI overview]]) consists of a 20 K focal plane unit hosting the corrugated feed horns, the orthomode transducers (OMTs) and the receiver front-end modules (FEMs). Forty-four composite waveguides {{BibCite|darcangelo2009a}} are interfaced with three conical thermal shields and connect the front-end modules to the warm (roughly 300 K) back-end unit (BEU) containing a further radio frequency amplification stage, detector diodes, and all the electronics for data acquisition and bias supply.
  
The Planck-LFI instrument is an array of 11 radiometric receivers in the Ka, Q and V bands, with centre frequencies close to 30, 44 and 70 GHz. The exact centre frequencies for each receiver are reported in <cite>#planck2011-1-4</cite> [{{PEarly|3|link=Planck early results. First assessment of the Low Frequency Instrument in-flight performance}}]; for simplicity, here we will refer to the three channels using their nominal centre frequency. A detailed description of the LFI instrument is given in <cite>#bersanelli2010</cite> [{{PPreLaunch|4|link=Planck pre-launch status: Design and description of the Low Frequency Instrument}}], and  references therein.
+
Best LFI noise performance is obtained with receivers based on InP High Electron Mobility Transistor (HEMT) low noise amplifiers (LNAs) for minimal power dissipation and best performance. To further minimize power consumption in the focal plane, the radiometers are split into two sub-assemblies connected by waveguides, one located at the telescope focal area, the other on the 300 K portion of the Planck satellite. These design features allow the entire front-end LNA dissipation to be <0.55 W, which enables the active cooling of the focal assembly. This is achieved with a vibration-less hydrogen sorption cooler, which also provides 18 K pre-cooling to the HFI helium J-T cooler. Two sorption cooler units are included in the flight hardware.
  
== The LFI description ==
+
As shown schematically in Fig. 1 below, the LFI consists of the following subsystems:
 +
* Radiometer Array Assembly (RAA);
 +
* Sorption Cooler Subsystem (SCS);
 +
* Radiometer Electronics Box Assembly (REBA).
  
The LFI instrument (see Figure 1 below) consists of a 20 K focal plane unit hosting the corrugated feed horns, the orthomode transducers (OMTs) and the receiver front-end modules (FEMs). Fourty four composite waveguides [O. D’Arcangelo, L. Figini, A. Simonetto et al. The Planck-LFI flight model composite waveguides. Journal of Instrumentation, 4:2007, 2009] are interfaced with three conical thermal shields and connect the front-end modules to the warm (~300 K) back-end unit (BEU) containing a further radio frequency amplification stage, detector diodes and all the electronics for data acquisition and bias supply.
+
The RAA includes the Front End Unit (FEU) and the Back End Unit (BEU), connected via waveguides. The FEU is located at the focus of the telescope, as one component of the joint LFI/HFI focal assembly (see sections below). The BEU is mounted on the top of the Planck service module (SVM).
 
+
The REBA (Radiometer Electronics Box Assembly) and the warm parts of the Sorption Cooler System (SCS) are located on one of the lateral panels of the SVM. The FEU and the Sorption Cooler Compressor (SCC) are connected by concentric stainless steel tubes. The smaller tube carries hydrogen at approximately 60 atmospheres from the cooler compressors to the FEU, while the larger tube returns the hydrogen at about 0.3 atmospheres. These units are described in following sections and in the [[LFIAppendix|Annexes]], while the SCS is described in details in the [[LFI design, qualification, and performance#The Sorption Cooler|Sorption Cooler]] section.
[[File:lfi_instrument.jpg|thumb|center|600px|Figure 1. Left Panel: the LFI instrument with main thermal stages, focal plane, waveguides and sorption cooler piping highlighted. Right Panel: Labelling of feed horns on the LFI focal plane.]]
 
 
 
Best LFI noise performance is obtained with receivers based on InP High Electron Mobility Transistor (HEMT) low noise amplifiers (LNAs) for minimal power dissipation and best performance. To further minimise power consumption in the focal plane, the radiometers are split into two sub-assemblies connected by waveguides, one located at the telescope focal area, the other on the 300 K portion of the Planck satellite. These design features allow the entire front-end LNAs dissipation to be <0.55 W, which enables the active cooling of the focal assembly. This is achieved with a vibration-less hydrogen sorption cooler, which also provides 18 K pre-cooling to the HFI helium J-T cooler. Two sorption cooler units are included in the flight hardware.
 
 
 
As shown schematically in Figure 2 below, the LFI consists of the following subsystems:
 
* Radiometer Array Assembly (RAA)
 
* Sorption Cooler Subsystem (SCS)
 
* Radiometer Electronics Box Assembly (REBA)
 
 
 
The RAA includes the Front End Unit (FEU) and the Back End Unit (BEU), connected via waveguides. The FEU is located at the focus of the telescope, as one component of the joint LFI/HFI focal assembly (see sections below). The BEU is mounted on the top of the Planck SVM.
 
The Radiometer Electronics Box Assembly (REBA) and the warm parts of the Sorption Cooler System (SCS) is located on one of the lateral panels of the service module (SVM). The FEU and the Sorption Cooler Compressor (SCC) are connected by concentric stainless steel tubes. The smaller tube carries hydrogen at ∼60 atmospheres from the cooler compressors to the FEU, while the larger tube returns the hydrogen at ∼0.3 atmospheres. These units are described in detail in following sections, the SCS is described in details in the satellite section dedicated to ''Cryogenics''.
 
 
All LFI units are linked together by the LFI harness, which also connects to the spacecraft interface.
 
All LFI units are linked together by the LFI harness, which also connects to the spacecraft interface.
  
[[File:schema.jpg|thumb|center|400px|Figure 2. Block Diagram of LFI]]
+
[[File:schema.jpg|thumb|center|500px|'''Figure 1. Block Diagram of LFI.''']]
 
 
 
 
  
 
=== ''Radiometer Array Assembly (RAA)'' ===
 
=== ''Radiometer Array Assembly (RAA)'' ===
  
 
The Radiometer Array Assembly (RAA) consists of two main units (the front end unit, FPU and the Back End Unit, BEU), connected by a set of waveguides.
 
The Radiometer Array Assembly (RAA) consists of two main units (the front end unit, FPU and the Back End Unit, BEU), connected by a set of waveguides.
The Focal Plane Unit (FPU) is the heart of the LFI instrument and it contains the feed array and associated orthomode transducers (OMTs) and FEMs, all cooled to 20 K by the sorption cooler. The FPU comprises a set of 11 modules, which are mounted on a mechanical support which meets the thermo-mechanical requirements of the instrument and adds thermal inertia.
+
The Focal Plane Unit (FPU) is the heart of the LFI instrument; it contains the feed array and associated orthomode transducers (OMTs) and FEMs, all cooled to 20 K by the sorption cooler. The FPU comprises a set of 11 modules, which are mounted on a mechanical support which meets the thermo-mechanical requirements of the instrument and adds thermal inertia.
 
The BEU comprises the radiometer Back End Modules (BEM) and the Data Acquisition Electronics (DAE), which are connected by an internal harness.
 
The BEU comprises the radiometer Back End Modules (BEM) and the Data Acquisition Electronics (DAE), which are connected by an internal harness.
 
The HFI Unit is located inside the LFI FPU and supported by the LFI structure. The LFI structure gives the mechanical and thermal interface to the HFI unit with the proper stiffness and thermal de-coupling. The LFI structure also guarantees the proper alignment of the HFI detector with the telescope focal plane.
 
The HFI Unit is located inside the LFI FPU and supported by the LFI structure. The LFI structure gives the mechanical and thermal interface to the HFI unit with the proper stiffness and thermal de-coupling. The LFI structure also guarantees the proper alignment of the HFI detector with the telescope focal plane.
  
The time scale of the stability of the receiver is driven by the 1 rpm rotation speed of the spacecraft, which requires a very low 1/f-noise or gain variation of the low noise amplifiers and other components.
+
The timescale of the stability of the receiverS is driven by the 1 rpm rotation speed of the spacecraft, which requires a very low 1/<i>f</i>-noise or gain variation of the low noise amplifiers and other components.
The LFI uses a pseudo-correlation receiver concept (Figure 3 below). This radiometer concept is chosen to maximise the stability of the instrument by reducing the effect of non-white noise generated in the radiometer itself. In this scheme, the difference between the inputs to each of the chains (the signal from the telescope and that from a reference black body respectively) is continuously being observed. To remove the effect of instability in the back-end amplifiers and detector diodes, it is necessary to switch the signal detected at the diodes at high rate.
+
The LFI uses a pseudo-correlation receiver concept (Fig. 2 below). This radiometer concept is chosen to maximize the stability of the instrument by reducing the effect of non-white noise generated in the radiometer itself. In this scheme, the difference between the inputs to each of the chains (the signal from the telescope and that from a reference blackbody, respectively) is continuously being observed. To remove the effect of instability in the back-end amplifiers and detector diodes, it is necessary to switch the signal detected at the diodes at a high rate.
 
The signals from the sky and from a reference load are combined by a hybrid coupler, amplified in two independent amplifier chains, and separated out by another hybrid. The sky and the reference load power can then be measured and differenced. Since the reference signal has been subject to the same gain variations in the two amplifier chains as the sky signal, the true sky power can be recovered.
 
The signals from the sky and from a reference load are combined by a hybrid coupler, amplified in two independent amplifier chains, and separated out by another hybrid. The sky and the reference load power can then be measured and differenced. Since the reference signal has been subject to the same gain variations in the two amplifier chains as the sky signal, the true sky power can be recovered.
 
The differencing receiver greatly improves the stability if the two input signals are almost equal, at
 
The differencing receiver greatly improves the stability if the two input signals are almost equal, at
a cost of a factor of <math>\sqrt 2 \; </math> in sensitivity compared to a perfectly stable total-power radiometer with the same noise temperature and bandwidth. This radiometer concept is capable of greatly reducing the knee frequency.
+
a cost of a factor of &radic;2 in sensitivity, compared to a perfectly stable total-power radiometer with the same noise temperature and bandwidth. This radiometer concept is also capable of greatly reducing the knee frequency.
We define as Radiometer Chain Assembly (RCA, see Figure 3) each functional unit from the feed horn to the BEM. The RAA therefore includes a set of 11 RCAs and the Data Acquisition Electronics, all mounted on a suitable mechanical structure (see also Figure 2 above). Although there are differences in the details of the radiometer chains at different frequencies, their overall configuration is similar, and a general description of its design is provided in this section.
+
A single Radiometer Chain Assembly (RCA; see Fig. 2) consists of each functional unit from the feed horn to the BEM. The RAA therefore includes a set of 11 RCAs and the Data Acquisition Electronics (see also Fig. 1 above), all mounted on a suitable mechanical structure. Although there are differences in the details of the radiometer chains at different frequencies, their overall configuration is similar, and a general description of their design is provided in this section.
Planck LFI has 11 Radiometer Chain Assembly (RCA). Each RCA is constituted by feed horn and FEM in the FEU (at 20 K), BEM (at 300 K) in the BEU and four waveguides that connect each FEM-BEM couple. The frequency distribution of the RCA is the following:
+
Planck LFI has 11 Radiometer Chain Assemblies (RCAs). Each RCA is constituted by a feed horn and FEM in the FEU (at 20 K), BEM (at 300 K) in the BEU, and four waveguides that connect each FEM-BEM couple. The frequency distribution of the RCA is the following:
* 2 RCAs at 30GHz
+
* 2 RCAs at 30 GHz;
* 3 RCAs at 44GHz
+
* 3 RCAs at 44 GHz;
* 6 RCAs at 70GHz
+
* 6 RCAs at 70 GHz.
  
[[File:rca_schematic.jpg|thumb|center|500px|Figure 3. A complete RCA from feed-horn to analog voltage output. The insets show the OMT, the details of the 20 K pseudo-correlator and of the back-end radio-frequency amplification, low-pass filtering, detection and DC amplification.]]
+
[[File:rca_schematic.jpg|thumb|center|640px|'''Figure 2. A complete RCA from feed-horn to analogue voltage output. The insets show the OMT, the details of the 20 K pseudo-correlator and of the back-end radio-frequency amplification, low-pass filtering, detection, and DC amplification.''']]
  
: ''Radiometer Chain Assembly (RCA)'';
+
==== ''Radiometer Chain Assembly (RCA)''====
  
RCAs consist of two radiometers, each feeding two diode detectors (see Figure 3 above), for a total of 44 detectors. The 11 RCAs are labelled by a numbers from 18 to 28 as outlined in Figure 1, right panel.
+
Every RCA consists of two radiometers, each feeding two diode detectors (see Fig. 2 above), for a total of 44 detectors. The 11 RCAs are labelled by a number from 18 to 28, as outlined in Fig. 1 of the [[LFI overview]], right panel.
  
Figure 3 provides a more detailed description of each radiometric receiver. In each RCA, the two perpendicular linear polarisation components split by the OMT propagate through two independent pseudo-correlation differential radiometers, labelled as ''M'' or ''S'' depending on the arm of the OMT they are connected to (''Main'' or ''Side'', see lower-left inset of Figure 3).
+
Fig. 2 provides a more detailed description of each radiometric receiver. In each RCA, the two perpendicular linear polarization components split by the OMT propagate through two independent pseudo-correlation differential radiometers, labelled as ''M'' or ''S,'' depending on the arm of the OMT they are connected to (''Main'' or ''Side'', see lower-left inset of Fig. 2).
  
In each radiometer the sky signal coming from the OMT output is continuously compared with a stable 4 K blackbody reference load mounted on the external shield of the HFI 4 K box [L. Valenziano, F. Cuttaia, A. De Rosa et al. Planck-LFI: design and performance of the 4 Kelvin Reference Load Unit. Journal of Instrumentation, 4:2006, 2009.]. After being summed by a first hybrid coupler, the two signals are amplified by ~30 dB, see upper-left inset of Figure 3. The amplifiers were selected for best operation at low drain voltages and for gain and phase match between paired radiometer legs, which is crucial for good balance. Each amplifier is labelled with codes ''1'', ''2'' so that the four outputs of the LNAs can be named with the sequence: ''M1'', ''M2'' (radiometer ''M'') and ''S1'', ''S2'' (radiometer ''S'').  
+
In each radiometer the sky signal coming from the OMT output is continuously compared with a stable 4 K blackbody reference load mounted on the external shield of the HFI 4-K box (Valenziano et al. 2009{{BibCite|valenziano2009}}). After being summed by a first hybrid coupler, the two signals are amplified by approximately 30 dB (see upper-left inset of Fig. 2). The amplifiers were selected for best operation at low drain voltages and for gain and phase match between paired radiometer legs, which is crucial for good balance. Each amplifier is labelled with codes ''1'' or ''2'' so that the four outputs of the LNAs can be named with the sequence: ''M1'', ''M2'' (radiometer ''M'') and ''S1'', ''S2'' (radiometer ''S'').  
Tight mass and power constraints called for a simple design of the Data Acquisition Electronics (DAE) box so that power bias lines were divided into five common-grounded power groups with no bias voltage readouts; only the total drain current flowing through the front-end amplifiers is measured and is available to the house-keeping telemetry (this design has important implications for front-end bias tuning, which depends critically on the satellite electrical and thermal configuration and was repeated at all integration stages, during on-ground and in-flight satellite tests).
+
Tight mass and power constraints called for a simple design of the Data Acquisition Electronics (DAE) box so that power bias lines were divided into five common-grounded power groups with no bias voltage readouts; only the total drain current flowing through the front-end amplifiers is measured and is available to the house-keeping telemetry (this design has important implications for front-end bias tuning, which depends critically on the satellite electrical and thermal configuration and was repeated at all integration stages during on-ground and in-flight satellite tests).
A phase shift alternating between <math> 0^\circ \; </math> and <math>180^\circ \; </math> at the frequency of 4096 Hz is applied in one of the two amplification chains and then a second hybrid coupler separates back the sky and reference load components that are further amplified and detected in the warm BEU, with a voltage output ranging from -2.5 V to +2.5 V.
+
A phase shift (or phase switch) alternating between 0&deg; and 180&deg; at the frequency of 4096 Hz is applied in one of the two amplification chains and then a second hybrid coupler restores the sky and reference load components, which are further amplified and detected in the warm BEU, with a voltage output ranging from -2.5 V to +2.5 V.
  
Each radiometer has two output diodes which are labelled with binary codes ''00'', ''01'' (radiometer ''M'') and ''10'', ''11'' (radiometer ''S''), so that the four outputs of each radiometric chain can be named with the sequence: ''M-00'', ''M-01'', ''S-10'', ''S-11''.
+
Each radiometer has two output diodes which are labelled with binary codes ''00'', ''01'' (radiometer ''M'') and ''10'', ''11'' (radiometer ''S''), so that the four outputs of each radiometric chain can be named with the sequence ''M-00'', ''M-01'', ''S-10'', ''S-11''.
  
After detection, an analog circuit in the DAE box removes a programmable offset in order to obtain a nearly null DC output voltage and a programmable gain is applied to increase the signal dynamics and optimally exploit the ADC input range. After the ADC, data are digitally downsampled, requantised and compressed in the radiometer electronics box assembly (REBA) according to a scheme described in [J. M. Herreros, M. F. Gómez, R. Rebolo et al. The Planck-LFI Radiometer Electronics Box Assembly. Journal of Instrumentation, 4:2008, 2009; M. Maris, M. Tomasi, S. Galeotta et al. Optimization of Planck-LFI on-board data handling. Journal of Instrumentation, 4:2018, 2009.] before preparing telemetry packets. On ground, telemetry packets are converted to sky and reference load time ordered data after calibrating the ADU samples into volt considering the applied offset and gain factors.
+
After detection, an analogue circuit in the DAE box removes a programmable offset in order to obtain a nearly null DC output voltage and a programmable gain is applied to increase the signal dynamics and optimally exploit the ADC input range. After the ADC, data are digitally down-sampled, re-quantized and compressed in the REBA according to a scheme described in (Herreros et al. 2009{{BibCite|herreros2009}}, Maris et al. 2009{{BibCite|maris2009}}), before being combined in telemetry packets. On the ground, telemetry packets are converted to time ordered data for both the sky and reference loadafter calibrating the ADU (Analog-to-Digital Unit) samples into volts.  The calibration takes into account the applied offset and gain factors.
  
To first order, the mean differential power output for each of the four receiver diodes can be written as follows [M. Seiffert, A. Mennella, C. Burigana et al. 1/f noise and other systematic effects in the Planck-LFI radiometers. A&A, 391:1185–1197, 2002; A. Mennella, M. Bersanelli, M. Seiffert et al. Offset balancing in pseudo-correlation radiometers for CMB measurements. A&A, 410:1089–1100, 2003; M. Bersanelli, N. Mandolesi, R. C. Butler et al. Planck pre-launch status: Design and description of the Low Frequency Instrument. A&A, 520:A4, 2010]:
+
To first order, the mean differential power output for each of the four receiver diodes can be written as follows (Seiffert et al. 2002{{BibCite|seiffert2002}}, Mennella et al. 2003{{BibCite|mennella2003}}, {{PlanckPapers|bersanelli2010}}):
  
: <math>
+
<math> \label{eq:power}
 
         P_{\rm out}^{\rm diode} = a\, G_{\rm tot}\,k\,\beta \left[ T_{\rm sky} + T_{\rm noise} - r\left(
 
         P_{\rm out}^{\rm diode} = a\, G_{\rm tot}\,k\,\beta \left[ T_{\rm sky} + T_{\rm noise} - r\left(
 
             T_{\rm ref} + T_{\rm noise}\right) \right]
 
             T_{\rm ref} + T_{\rm noise}\right) \right]
 
</math>
 
</math>
  
where <math>G_{\rm tot}</math> is the total gain, <math>k \;</math> is the Boltzmann constant, <math>\beta</math> the receiver bandwidth and <math>a</math> is the diode constant. <math>T_{\rm sky} \;</math> and <math>T_{\rm ref}\; </math> are the average sky and reference load antenna temperatures at the inputs of the first hybrid and <math>T_{\rm noise}</math> is the receiver noise temperature.
+
where <i>G</i><sub>tot</sub> is the total gain, <i>k</i> is the Boltzmann constant, <i>&beta;</i> the receiver bandwidth, and <i>a</i> is the diode constant. <i>T</i><sub>sky</sub> and <i>T</i><sub>ref</sub> are the average sky and reference load antenna temperatures at the inputs of the first hybrid and <i>T</i><sub>noise</sub> is the receiver noise temperature.
 +
 
 +
The gain modulation factor (Mennella et al. 2003{{BibCite|mennella2003}}, {{PlanckPapers|planck2011-1-6 }}), <i>r</i>, is defined by
  
The gain modulation factor [A. Mennella, M. Bersanelli, M. Seiffert et al. Offset balancing in pseudo-correlation radiometers for CMB measurements. A&A, 410:1089–1100, 2003; A. Zacchei, D. Maino, C. Baccigalupi et al. Planck early results. V. The Low Frequency Instrument data processing. A&A, 536:A5, 2011], <math>r</math>, is defined by:
+
<math> \label{eq:erre1}
: <math>
+
         r = \frac{T_{\rm sky} + T_{\rm noise}}{T_{\rm ref} + T_{\rm noise}} ,
         r = \frac{T_{\rm sky} + T_{\rm noise}}{T_{\rm ref} + T_{\rm noise}}  
 
 
</math>
 
</math>
  
and is used to balance (in software) the temperature offset between the sky and reference load signals and minimise the residual 1/<math>f</math> noise in the differential datastream. This parameter is calculated from the average uncalibrated total power data using the relationship:
+
and is used to balance (in software) the temperature offset between the sky and reference load signals and minimize the residual 1/<i>f</i> noise in the differential data stream. This parameter is calculated from the average uncalibrated total power data using the relationship
: <math>
+
 
 +
<math> \label{eq:erre2}
 
         r = \langle V_{\rm sky} \rangle/ \langle V_{\rm ref}\rangle,
 
         r = \langle V_{\rm sky} \rangle/ \langle V_{\rm ref}\rangle,
 
</math>
 
</math>
  
where <math><V_{\rm sky}></math> and <math><V_{\rm ref}></math> are the average sky and reference voltages calculated in a defined time range.
+
where <<i>V</i><sub>sky</sub>> and <<i>V</i><sub>ref</sub>> are the average sky and reference voltages calculated in a defined time range.
The white noise spectral density at the output of each diode is essentially independent from the reference-load absolute temperature and is given by:
+
The white noise spectral density at the output of each diode is essentially independent of the reference-load absolute temperature and is given by
: <math>
+
 
 +
<math> \label{eq:dt1}
 
         \Delta T_0^{\rm diode} = \frac{2\,(T_{\rm sky}+T_{\rm noise})}{\sqrt{\beta}}.
 
         \Delta T_0^{\rm diode} = \frac{2\,(T_{\rm sky}+T_{\rm noise})}{\sqrt{\beta}}.
 
</math>
 
</math>
  
 +
If the front-end components are not perfectly balanced, then the separation of the sky and reference load signals after the second hybrid is not perfect and the outputs are mixed. First-order deviations in white noise sensitivity from the ideal behaviour are caused mainly by noise temperature and phase-switch amplitude mismatches. Following the notation used in (Seiffert et al. 2002 {{BibCite|seiffert2002}}), we define &epsilon;<sub><i>T</i><sub>n</sub></sub>, as the imbalance in front end noise temperature, and &epsilon;<sub><i>A</i><sub>1</sub></sub> and &epsilon;<sub><i>A</i><sub>2</sub></sub>, the imbalance in signal attenuation in the two states of the phase switch. Equation above for the two diodes of a slightly imbalanced radiometer then becomes
 +
 +
<math> \label{eq:dt2} (\Delta T^{\rm diode} )^2 ≈ (\Delta T_0^{\rm diode})^2 ( 1± \frac{\epsilon_{A1}- \epsilon_{A2}}{2}
 +
+ \alpha \epsilon_{T{\rm n}}),
 +
</math>
 +
 +
which is identical for the two diodes apart from the sign of the term &epsilon;<sub><i>A</i><sub>1</sub></sub> − &epsilon;<sub><i>A</i><sub>2</sub></sub>, representing the phase switch amplitude imbalance. This indicates that the isolation loss caused by this imbalance generates an anti-correlation between the white noise levels of the single-diode data streams.
 +
For this reason, the LFI scientific data streams are obtained by averaging the voltage outputs from the two diodes in each radiometer:
 +
 +
<math> \label{eq:v1}
 +
V^{\rm rad}_{\rm out} = w_1 V^{\rm diode\,1}_{\rm out} +w_2 V^{\rm diode\,2}_{\rm out} </math>,
 +
 +
where &omega;<sub>1</sub> and &omega;<sub>2</sub> are inverse-variance weights calculated from the data as discussed in ({{PlanckPapers|planck2011-1-6}}). This way, the diode-diode anti-correlation is cancelled, and the radiometer white noise becomes
 +
 +
<math> \label{eq:dt3}
 +
\Delta T^{\rm rad} ≈ \frac{ T_0^{\rm diode}}{\sqrt{2}} (1+\alpha \epsilon_{T{\rm n}})^{1/2}.
 +
</math>
 +
 +
In the equations above, &epsilon;&laquo;1, while &alpha;  (a term &asymp;1) is given by
 +
 +
<math> \label{eq:alpha}
 +
\alpha = \frac{ T_{\rm noise} (2 \; T_{\rm noise} + T_{\rm sky} + T_{\rm ref} }
 +
{2 (T_{\rm sky} + T_{\rm noise} )(T_{\rm ref} + T_{\rm noise})}.
 +
</math>
 +
 +
See the [[TOI processing LFI|Diode Combination]] section for the details of the diode combination procedure.
  
In Figure 4 below we show a close-up of the two front end modules of an RCA with the four phase switches which are labelled with the four letters ''A'' and ''B'' (main arm), ''C'' and ''D'' (side arm). Each phase switch is characterised by two states: state 0 (no phase shift applied to the incoming wave) and state 1 (<math>180^\circ</math> phase shift applied) and can either stay fixed in a state or switch at 4 kHz between the two states.
+
In Fig. 3 below we show a close-up of the two front end modules of an RCA with the four phase switches, which are labelled with the four letters ''A'' and ''B'' (main arm), and ''C'' and ''D'' (side arm). Each phase switch is characterized by two states: state 0 (no phase shift applied to the incoming wave); and state 1 (180&deg; phase shift applied) and can either stay fixed in a state or switch at 4 kHz between the two states.
 
      
 
      
Phase switches are clocked and biased by the DAE and their configuration can be programmed via telecommand. In order to simplify the instrument electronics, phase switches are configured and operated in pairs: ''A/C'' and ''B/D''. This means that if phase switches ''A'' and ''C'' are switching at 4 kHz then ''B'' and ''D'' are fixed both in the same state (either 0 or 1). This simplification, required during the design phase to comply with mass and power budgets, comes at the price of loosing some setup redundancy.
+
Phase switches are clocked and biased by the DAE and their configuration can be programmed via telecommand. In order to simplify the instrument electronics, phase switches are configured and operated in pairs, by convention they are labelled ''A/C'' (corresponding to the first LNA both of main and side arm) and ''B/D'' (corresponding to the second LNA both of main and side arm). This means that if phase switches ''A'' and ''C'' are switching at 4 kHz then ''B'' and ''D'' are fixed, both in the same state (either 0 or 1); and viceversa. This simplification, required during the design phase to comply with mass and power budgets, comes at the price of losing some setup redundancy.
 +
 
 +
[[File:phase_switch_operation.jpg|thumb|center|460px|'''Figure 3. Close-up of the two front-end modules of an RCA. There are four phase switches, labelled A, B, C, and D. Each switch can be fixed in one of its two positions (labelled as 0, 1) or switched at 4 kHz between 0 and 1. Phase switches are clocked and biased by the DAE in pairs: A/C and B/D.''']]
 +
 
 +
==== ''Feed Horns (FHs)''====
  
[[File:phase_switch_operation.jpg|thumb|center|400px|Figure 4. Close-up of the two front-end modules of an RCA. There are four phase switches, labelled A, B, C and D. Each switch can be fixed in one of its two positions (labelled as 0, 1) or switch at 4 kHz between 0 and 1. Phase switches are clocked and biased by the DAE in pairs: A/C and B/D (see text for further explanation).]]
+
Dual profiled corrugated horns have been selected at all LFI frequencies as the best design in terms of the shape of the main lobe, level of the side lobes, control of the phase centre, and compactness.  
 +
Details of the design, flight model and tests of Planck-LFI feed horns can be found in (Villa et al. 2009{{BibCite|villa2009}}) and in the corresponding [[LFIAppendix#Feed Horns (FH)|Annex]] section.
  
: ''Feed Horns (FH)''
+
==== ''Ortho-Mode Transducers (OMTs)''====
  
Dual profiled corrugated horns have been selected at all LFI frequencies as the best design in terms of shape of the main lobe, level of the side lobes, control of the phase centre, and compactness. Dual profiled LFI horns are composed by a Sine Squared profiled section, and an exponential profile near the aperture plane.  
+
The Ortho–Mode Transducer (OMTs) separates the radiation collected by the feedhorn into two orthogonal polarization components. It consists of a circular to square waveguide transition (directly connected to the FH), a square waveguide section and two separate rectangular waveguides (the main and side arms, which separate and pick up the orthogonal polarizations, connected with the FEU). A 90&deg; bend is always present on the side arm, while a twist is also necessary on the main (30 and 44 GHz) and side (70 GHz) arms, in order to match the FEU polarization.
In order to optimise the optical matching of the feeds phase centres to the telescope focal surface, while preventing obscuration between horns, LFI has 6 different feed horn designs. For each frequency, the number of feeds and the number of different designs are reported in Table 1 below
 
  
{| border="1" cellspacing="0" cellpadding="2" align="center"
+
The details of the flight models and measurements of the Planck LFI ortho-mode transducers can be found in (D'Arcangelo et al. 2009{{BibCite|darcangelo2009b}}) and in the corresponding [[LFIAppendix#OrthoMode Transducers (OMT)|Annex]] section.
|+ '''Table 1 - Number of LFI feed horns and number of different feed horn designs'''
+
 
|-
+
==== ''Front End Modules (FEMs)''====
!scope="col"| Frequency (GHz)
+
 
!scope="col"| Number of Horns
+
Front End Modules are located in the FPU, just behind the Feed Horn and the Ortho Mode Transducers. The 70 GHz FEMs are mounted onto the inner wall of the main frame (the wall facing HFI instrument) from the HFI side. The 44 and 30 GHz FEMs are inserted into the main frame from the waveguide (WG) side and fixed to the bottom plate. Screws to the bottom plate are inserted from the WG side.
!scope="col"| Number of designs
+
The LFI FEMs are the first active stage of amplification of the radiometer chain. Each FEM contains four amplification paths, each of which is composed of several cascaded LNAs followed by a phase switch. Two passive hybrids, at the input and output of the FEM, are used to mix pairs of signals of the same radiometer (see Fig. 3). This cuases the instabilities of each chain to be applied to both the sky and load signals.
|-
+
 
|width="120" |30
+
The passive hybrid coupler ("magic-tee") combines the signals from the sky and cold load with a fixed phase offset of either 90&deg; or 180&deg; between them. It has a 20% bandwidth, low loss, and amplitude balance needed at the output to ensure adequate signal isolation.
|width="120" |2
+
 
|width="120" |1
+
The details of the design, development and verification of the 70 GHz front-end modules for the Planck Low Frequency Instrument can be found in (Varis et al. 2009{{BibCite|varis2009}}) and in the corresponding [[LFIAppendix#Front End Modules (FEM)|Annex]] section.
|-
+
 
|width="120" |44
+
==== ''Waveguides (WGs)''====
|width="120" |3
+
 
|width="120" |2
+
The LFI Front End Unit (FEU) is connected to the Back End Unit (BEU) by 44 rectangular waveguides approximately 1.5-2.0 m long. Each waveguide exhibits a low voltage standing wave ratio, low thermal conductivity, low insertion loss, and low mass. In addition, the waveguide path permits the LFI/HFI integration and the electrical bonding between the FPU and the BEU. Because of the Focal Plane Unit arrangement, the waveguides are in general twisted and bent in different planes and with different angles, depending on the particular waveguide. From the thermal point of view the waveguides have to connect two systems (the BEM and FEM) that are at very different temperatures. At BEM level the waveguides are at a temperature of 300 K, while at FEM level the temperature is 20 K. The waveguides have to reduce the thermal flow from 300 K to 20 K. In Fig. 1 in the [[LFI overview]] (left panel) a conceptual sketch of the LFI configuration is shown.
|-
+
 
|width="120" |70
+
Details of the Planck-LFI flight model of the composite waveguides can be found in (D'Arcangelo et al. 2009{{BibCite|darcangelo2009a}}) and in the corresponding [[LFIAppendix#Waveguides|Annex]] section.
|width="120" |6
+
 
|width="120" |3
+
==== ''Back End Modules (BEMs)''====
|-
+
 
|}
+
The BEMs are composed of four identical channels each made of Low Noise Amplifiers (LNAs), an RF bandpass Filter, RF to DC diode detector, and DC amplifiers.
 +
The FEM output signals are connected by waveguides from the Focal Plane Unit (FPU) assembly to the Back End Modules (BEMs) housed adjacent to the Data Acquisition Electronics (DAE) assembly. To maintain compatibility with the FEMs, each BEM accommodates four receiver channels from the four waveguide outputs of each FEM. The BEM internal signal routes are not cross-coupled and can be regarded as four identical parallel circuits.
 +
Each BEM is constructed as two mirror halves. The two amplifier/detector assemblies each contain two amplifier/detector circuits. Each is supplied from one of a pair of printed circuit boards, which also house two DC output amplifiers.
 +
 
 +
The details of the design, development and verification of the 30 and 44 GHz back-end modules for the Planck Low Frequency Instrument can be found in (Artal et al. 2009{{BibCite|artal2009}}).
 +
Details of the design, development and verification of the 70 GHz back-end modules for the Planck Low Frequency Instrument can be found in (Varis et al. 2009{{BibCite|varis2009}}).
 +
Details are also reported in the corresponding [[LFIAppendix#Back End Modules (BEM)|Annex]] section.
 +
 
 +
==== ''4K Load'' ====
 +
 
 +
The purpose of the 4K reference load is to provide the radiometer with a stable reference signal. Reducing the input offset (the radiometric temperature difference between the sky and the reference load) reduces the minimum achievable radiometer 1/<i>f</i> noise knee frequency for a given amplifier fluctuation spectrum. A reference load temperature that matches the sky temperature (approximately 2.7 K) would be ideal.
 +
 
 +
Details of the design, characteristics, and performance of the LFI 4-K reference load units are given in (Valenziano et al. 2009{{BibCite|valenziano2009}}) and in the corresponding section of the [[LFIAppendix#4K Load|Annexes]].
 +
 
 +
=== Naming Convention ===
 +
 
 +
The naming of all the LFI elements has been described in the previous sections, but here is summarized again for clarity.
 +
 
 +
The 11 RCAs are labelled by numbers from 18 to 28, as outlined in Fig. 1 in the [[LFI overview]] (right panel). In each RCA, the two perpendicular linear polarization components are labelled as ''M'' or ''S'' according to the arm of the OMT they are connected to (''Main'' or ''Side'', see lower-left inset of Fig. 2).
 +
 
 +
Each front-end amplifier (see upper-left inset of Fig. 2) is labelled with the codes ''1'', ''2'', so that the four outputs of the FEM LNAs can be named with the sequence ''M1'', ''M2'' (radiometer ''M'') and ''S1'', ''S2'' (radiometer ''S'').
 +
 
 +
Each radiometer has two output diodes (see upper-right inset of Fig. 2), which are labelled with binary codes ''00'', ''01'' (radiometer ''M'') and ''10'', ''11'' (radiometer ''S''), so that the four outputs of each radiometric chain can be named with the sequence ''M-00'', ''M-01'', ''S-10'', and ''S-11''.
 +
 
 +
=== ''REBA'' ===
 +
 
 +
The Radiometer Electronics Box Assembly (REBA) is the electronics unit that processes the digitized scientific data and manages the overall instrument. It is also in charge of communication with the spacecraft.
 +
There are two REBA boxes, one nominal and one redundant. The redundancy concept is cold, which means that both boxes are never ON at the same time; the operation of each unit is managed by the spacecraft switching on the corresponding unit. The REBA ASW (Application SoftWare) is the same in each REBA box.
 +
 
 +
A detailed description of the Planck LFI REBA can be found in (Herreros et al. 2009{{BibCite|herreros2009}}) and in the corresponding section of the [[LFIAppendix#REBA|Annexes]].
 +
 
 +
=== ''Instrument On-board Software'' ===
 +
 
 +
The REBA software is the on-board software of LFI. It is installed in the two computing subunits of REBA: the DPU (Digital Processing Unit), responsible of the control and monitoring of the instrument and the interface with the spacecraft; and the SPU (Signal Processing Unit), responsible for the data reduction and compression.
 +
 
 +
Details can be found in the corresponding section of the [[LFIAppendix#Instrument On-board Software|Annexes]].
 +
 
 +
==== ''Reduction and Compression of Science Data'' ====
  
The Table 2 below reports the performances and characteristics of the LFI feed horns. Although the required return loss is 25 dB (including flange effects and fabrication tolerances), for the design process the return loss has been set at –30 dB. The Return Loss and the cross polar values reported in the table are the worst values over the entire 20% of bandwidth based on measurements.
+
To assess stability against 1/<i>f</i> noise, the Low Frequency Instrument (LFI) on board the Planck mission acquired data at a rate much higher than the data rate allowed by the science telemetry bandwidth of 35.5 kbps. The data were processed by an on-board pipeline, followed on the ground by a decoding and reconstruction step, to reduce the volume of data to a level compatible with the bandwidth, while minimizing the loss of information. The on-board processing of the scientific data used by Planck/LFI to fit the allowed data-rate is an intrinsically lossy process which distorts the signal in a manner that depends on a set of five free parameters (<i>N</i><sub>aver</sub>, <i>r</i><sub>1</sub>, <i,>r</i><sub>2</sub>, <i>q</i>, <i>O</i>) for each of the 44 LFI detectors.
 +
A brief description of the characteristics of this algorithm and the level of distortion introduced by the on-board processing as a function of these parameters can be found in the corresponding section of the [[LFIAppendix#Reduction and Compression of Science Data|Annexes]], while a full description of the Planck LFI on-board data handling system and the tuning and optimization method of the on-board processing chain can be found in (Maris et al. 2009{{BibCite|maris2009}}).
  
 +
The strategy adopted to fit into the telemetry bandwidth relies on three on-board processing steps: downsampling; pre-processing the data to ensure loss-less compression; and loss-less compression itself. To demonstrate these steps, a model of the input signal was used. Note that while the compression is loss-less, the pre-processing is not, due to the need to rescale the data and convert them to integers (a process named data re-quantization). However, the whole strategy is designed to keep strict control over the way in which lossy operations are done, and to quantify the amount of information loss in order to assess the optimal compression rate with minimal information loss.
  
{| border="1" cellspacing="0" cellpadding="2" align="center"
+
=== ''Instrument Operations'' ===
|+ '''Table 2 - Performance and characteristics of the LFI feed horns as derived from measurements at 300 K'''
 
|-
 
!scope="col"| Horn ID
 
!scope="col"| Bandwidth [GHz]
 
!scope="col"| Edge Taper [dB@22°]
 
!scope="col"| X-pol Centre frequency [dB]
 
!scope="col"| Return–Loss [dB]
 
!scope="col"| Side Lobes Centre frequency [dB]
 
|-
 
|width="80" |18
 
|width="120" |14
 
|width="120" |17.6
 
|width="120" |-34
 
|width="120" |-38
 
|width="120" |-40
 
|-
 
|width="80" |19
 
|width="120" |14
 
|width="120" |17.0
 
|width="120" |-32
 
|width="120" |-34
 
|width="120" |-36
 
|-
 
|width="80" |20
 
|width="120" |14
 
|width="120" |17.4
 
|width="120" |-34
 
|width="120" |-35
 
|width="120" |-36
 
|-
 
|width="80" |21
 
|width="120" |14
 
|width="120" |17.4
 
|width="120" |-36
 
|width="120" |-34
 
|width="120" |-36
 
|-
 
|width="80" |22
 
|width="120" |14
 
|width="120" |17.5
 
|width="120" |-32
 
|width="120" |-34
 
|width="120" |-36
 
|-
 
|width="80" |23
 
|width="120" |14
 
|width="120" |17.4
 
|width="120" |-41
 
|width="120" |-34
 
|width="120" |-39
 
|-
 
|width="80" |24
 
|width="120" |8.8
 
|width="120" |31.5
 
|width="120" |-34
 
|width="120" |-30
 
|width="120" |-42
 
|-
 
|width="80" |25
 
|width="120" |8.8
 
|width="120" |31.5
 
|width="120" |-36
 
|width="120" |-31
 
|width="120" |-36
 
|-
 
|width="80" |26
 
|width="120" |8.8
 
|width="120" |32.0
 
|width="120" |-35
 
|width="120" |-31
 
|width="120" |-36
 
|-
 
|width="80" |27
 
|width="120" |6
 
|width="120" |30.3
 
|width="120" |-42
 
|width="120" |-28
 
|width="120" |-34
 
|-
 
|width="80" |28
 
|width="120" |6
 
|width="120" |30.0
 
|width="120" |-42
 
|width="120" |-28
 
|width="120" |-34
 
|-
 
|}
 
  
 +
==== ''Operational Modes'' ====
  
 +
The operations of the LFI are designed to be automatic and require little if any intervention from the ground. A small number of commands are required for operating the instrument and eventually for diagnostic and reconfiguration purposes.
 +
Each sky survey is conducted by the LFI with the instrument in the Normal Operations Mode. No deployable elements, or mechanically moving parts are included in the instrument. The scanning of the sky is achieved by progressive repointing of the satellite spin axis, with the Sun direction always within a cone 10&deg; from the spin axis.
 +
Within the Normal Science Mode the instrument can be configured in order to fit with different science or diagnostic needs without changing the power consumption and thus the temperature in the FPU.
 +
<!--
 +
Changes in power consumption in the FPU are minimized and should occur only in the case that failures in the radiometers that could create interference problems require an RCA to be switched off. Power adjustments on the first stage of the HEMT amplifiers (which were contemplated), require extremely small power level variations.-->
  
: ''OrthoMode Transducers (OMT)''
+
A brief summary of the LFI Operational Modes and the transitions between them is given in the corresponding section of the [[LFIAppendix#LFI Operational Modes|Annexes]].
  
The Ortho–Mode Transducer (OMTs) separates the two orthogonal polarisation components propagating inside the horn.
+
<!--
The OMT is composed by the central square waveguide with two slots (the through and the side one, which separate and pick up the two polarisations), the waveguide transition between the two slots and the WR28 standard waveguide, the 90 degrees bend for the side port, the 90 degrees stepped twist for the through port waveguide transition between the square and the input circular waveguide of the horn.
+
==== ''In-flight Operations'' ====
One polarisation propagates directly from the circular waveguide to the rectangular on–axis waveguide. A circular to rectangular stepped transition provides the interface with a WR28 standard waveguide, and the short for the other orthogonal polarisation. The second polarisation component propagates in the side arm.
 
  
The performances for the LFI OMTs at all frequencies are reported in the following table:
+
TBW
  
{| border="1" cellspacing="0" cellpadding="2" align="center"
 
|+ '''Table 3 Performance Characteristics of the LFI OMTs based on measurements. The values are the worst values over the entire 20% of bandwidth expect the Insertion Loss that is the average value'''
 
|-
 
!scope="col"| OMT ID
 
!scope="col"| Bandwidth [GHz]
 
!scope="col"| X–Pol [dB] (Main)
 
!scope="col"| X–Pol [dB] (Side)
 
!scope="col"| Insertion–Loss [dB] (Main)
 
!scope="col"| Insertion–Loss [dB] (Side)
 
!scope="col"| Return–Loss [dB] (Main)
 
!scope="col"| Return–Loss [dB] (Side)
 
|-
 
|width="80" |18
 
|width="100" |14
 
|width="100" |<29
 
|width="100" |<30
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |19
 
|width="100" |14
 
|width="100" |<26
 
|width="100" |<28
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |20
 
|width="100" |14
 
|width="100" |<32
 
|width="100" |<35
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |21
 
|width="100" |14
 
|width="100" |<32
 
|width="100" |<37
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |22
 
|width="100" |14
 
|width="100" |<26
 
|width="100" |<28
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |23
 
|width="100" |14
 
|width="100" |<26
 
|width="100" |<28
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-15.0
 
|width="100" |-18.0
 
|-
 
|width="80" |24
 
|width="100" |8.8
 
|width="100" |<38
 
|width="100" |<40
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-13.0
 
|width="100" |-15.0
 
|-
 
|width="80" |25
 
|width="100" |8.8
 
|width="100" |<31
 
|width="100" |<32
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-13.0
 
|width="100" |-15.0
 
|-
 
|width="80" |26
 
|width="100" |8.8
 
|width="100" |<27
 
|width="100" |<25
 
|width="100" |0.15
 
|width="100" |0.2
 
|width="100" |-13.0
 
|width="100" |-15.0
 
|-
 
|width="80" |27
 
|width="100" |6
 
|width="100" |<38
 
|width="100" |<44
 
|width="100" |0.15
 
|width="100" |0.10
 
|width="100" |-16.0
 
|width="100" |-24.0
 
|-
 
|width="80" |28
 
|width="100" |6
 
|width="100" |<36
 
|width="100" |<38
 
|width="100" |0.15
 
|width="100" |0.15
 
|width="100" |-16.0
 
|width="100" |-24.0
 
|-
 
|}
 
  
 +
==== ''Anomalies'' ====
  
 +
TBW
 +
-->
  
: ''Front End Modules (FEM)''
+
==<span id="LFITests">Ground Tests</span>==
 +
During its development, the LFI flight model was calibrated and tested at various integration levels from sub-systems (Davis et al. 2009{{BibCite|davis2009}}) to individual integrated receivers ({{PlanckPapers|villa2010}}) and the whole receiver array ({{PlanckPapers|mennella2010}}). In every campaign we performed tests according to the following classification:
  
Front End Modules are located in the FPU, just in cascade of the Feed Horn and the Ortho Mode Transducers. 70 GHz FEMs are mounted onto the inner wall of the mainframe (the wall facing HFI instrument) from the HFI side. 44 and 30 GHz FEMs are inserted into the mainframe from the WG side and fixed to the bottom plate. Screws to bottom plate are inserted from WG side.
+
* ''functionality tests'', performed to verify the instrument functionality;
The LFI FEMs are the first active stage of amplification of the radiometer chain. Each FEM contains four amplification paths. Each path is composed by several cascaded LNAs followed by a phase switch. Two passive hybrids, at the input and output of the FEM, are used to mix couples of signals of the same radiometer (see Figure 4). This makes the instabilities of each chain to be applied to both the sky and load signals.
+
* ''tuning tests'', to tune radiometer parameters (biases, DC electronics gain and offset, digital quantization and compression) for optimal performance in flight-like thermal conditions;
 +
* ''basic calibration and noise performance tests'', to characterize instrument performance (photometric calibration, isolation, linearity, noise and stability) in tuned conditions;
 +
* ''susceptibility tests'', to characterize instrument susceptibility to thermal and electrical variations.
  
: ''Waveguides''
+
Where possible, the same tests were repeated in several test campaigns, in order to ensure enough redundancy and confidence in the repeatability of the instrument behaviour. A matrix showing the instrument parameters measured in the various test campaigns is provided in Table 1 of ({{PlanckPapers|mennella2010}}).
  
The LFI Front End Unit (FEU) (cooled at 20 K by the sorption cooler) is connected to the 300 K Back End Unit (BEU) by 44 rectangular waveguides approximately 2.0 meter long. Each waveguide exhibits low VSWR, low thermal conductivity, low insertion loss, and low mass. In addition, the waveguide path shall permit the LFI/HFI integration and the electrical bonding between FPU and BEU. Because of the Focal Plane Unit arrangement, the waveguides are in general twisted and bended in different planes and with different angles, depending on the particular waveguide. In Figure 1 (left panel) a conceptual sketch of the LFI configuration is shown.
+
The ground test campaign was developed in three main phases: cryogenic tests on the individual RCAs; cryogenic tests on the integrated receiver array (the so-called radiometer array assembly, RAA); and system-level tests after the integration of the LFI and HFI instruments onto the satellite. The first two phases were carried out at the Thales Alenia Space - Italia laboratories located in Vimodrone (Milano, Italy) (note that receiver tests on 70 GHz RCAs were carried out in Finland, at Yilinen laboratories), while system level tests (SLTs) were conducted in a dedicated cryofacility at the Centre Spatiale de Li&egrave;ge (CSL) located in Liege, Belgium.  
  
The performance for the LFI waveguides at all frequencies are reported in the following table:
+
In Table 1 below we list the temperature of the main cold thermal stages during ground tests compared to in-flight nominal values. These values show that system-level tests were conducted in conditions that were as flight-representative as possible, while results obtained during RCA and RAA tests need to be extrapolated to flight conditions to allow comparison. Details of the RCA test campaign are discussed in ({{PlanckPapers|villa2010}}), while the RAA tests and the extrapolation methods are presented in ({{PlanckPapers|mennella2010}}).
  
 
{| border="1" cellspacing="0" cellpadding="2" align="center"
 
{| border="1" cellspacing="0" cellpadding="2" align="center"
|+ '''Table 4 Number of LFI waveguides and performances. The Insertion Loss (IL), Return Loss(RL) and Electrical Resistance (R) values are the requirements. In between parenthesis the goal is reported also. Note that the RL value includes possible degradation due to presence of flanges.'''
+
|+ '''<small>Table 1. Temperatures of the main cold stages during the various ground test campaigns compared to in-flight nominal values.</small>'''
 
|-
 
|-
!scope="col"| Frequency
+
!scope="col"| Temperature
!scope="col"| Type
+
!scope="col"| Nominal
!scope="col"| Number
+
!scope="col"| RCA tests
!scope="col"| IL [dB]
+
!scope="col"| RAA tests
!scope="col"| RL [dB]
+
!scope="col"| System-level
!scope="col"| R [mΩ] @20 K
 
!scope="col"| R [mΩ] @300 K
 
 
|-
 
|-
|width="80" | 30
+
|width="120" | Sky
|width="100" |WR28
+
|width="100" | ~ 3 K
|width="100" |8
+
|width="100" | 8 K
|width="100" |2.0 (1.0)
+
|width="100" | ≳ 18.5 K
|width="100" |≤ −20
+
|width="100" | ~ 4 K
|width="100" |11.8
 
|width="100" |27.3
 
 
|-
 
|-
|width="80" | 44
+
|width="120" | Ref. load
|width="100" |WR22
+
|width="100" | ~ 4.5 K
|width="100" |12
+
|width="100" | ≳ 8 K
|width="100" |3.0 (1.5)
+
|width="100" | ≳ 18.5 K
|width="100" |≤ −20
+
|width="100" | ~ 4.5 K
|width="100" |14.7
 
|width="100" |34.1
 
 
|-
 
|-
|width="80" | 70
+
|width="120" | Front-end unit
|width="100" |WR12
+
|width="100" | ~ 20 K
|width="100" |24
+
|width="100" | ~ 20 K
|width="100" |4.0 (3.5)
+
|width="100" | ~ 26 K
|width="100" |≤ −20
+
|width="100" | ~ 20 K
|width="100" |26.2
 
|width="100" |60.5
 
 
|-
 
|-
 
|}
 
|}
  
 +
During the various test campaigns the instrument was switched off and moved several times in a time period of about three years. A series of functional tests were always repeated at each location and also in-flight, in order to verify the instrument functionality and the response repeatability. No failures or major problems have been identified due to transport and integration procedures.
 +
 +
The expected Planck LFI scientific performance, resulting mainly from cryogenic system level tests, are described in ({{PlanckPapers|mennella2010}}).
 +
 +
== <span id="LFICalibration">In-flight Calibration</span>==
 +
 +
The LFI Commissioning and Calibration and Performance Verification (CPV) phases started on 4 June 2009 and lasted until 12 August 2009 when Planck started scanning the sky in nominal mode. At the onset of CPV, the active cooling started when the radiating surfaces on the payload module reached their working temperatures (approximately 50 K on the third V-groove, and 40 K on the reflectors) by passive cooling. This was achieved during the transfer phase. Nominal temperatures were achieved on 3 July 2009, when the dilution cooler temperature reached 0.1 K ({{PlanckPapers|planck2011-1-3}}, Meinhold et al. 2009 {{BibCite| meinhold2009}}). The cooldown of the HFI 4-K stage (see Fig. 4 below), was key during CPV for the LFI, because it provided a variable input signal that was exploited during bias tuning.
 +
 +
The LFI Commissioning and CPV was carried out in four phases:
 +
* LFI switch-on and basic functionality verification (Commissioning);
 +
* tuning of front-end biases and back-end electronics (CPV);
 +
* preliminary calibration tests (CPV);
 +
* thermal tests (CPV).
 +
 +
[[File:FUNCT_tests_schedule_vs_K-eps-converted-to.jpg|thumb|center|460px|'''Figure 4. Functional tests performed during CPV against timeline. Curves from thermal sensors monitoring the 4-K stage and the FPU temperature are superimposed.''']]
  
All the required characteristics cannot be realised with single material waveguide configuration; a composite waveguide configuration is needed.  
+
Details of the LFI Commissioning and CPV test campaign are given in (Gregorio et al. 2013{{BibCite|gregorio2013}}).
  
From the thermal point of view the waveguides have to connect 2 systems (BEM and FEM) that are at different temperatures. At BEM level the waveguides are at a temperature of 300K while at FEM level the temperature is 20K. Along the Stainless Steel section the waveguides have to reduce the thermal flow from 300K to 20K. The Stainless Steel waveguide is connected to all the V-grooves in order to dissipate the heat produced at BEU level.
+
==<span id="LFIPerformance">Performance Summary</span>==
The copper waveguides section is connected to a mechanical support structure in five points in order to increase the stiffness of the waveguide.
 
  
Each copper waveguide has different path and, whereas the straight Stainless Steel sections are identical for all the channels except for the cross dimensions that define the cut-off frequency.
+
A summary of the LFI performance parameters is given in Table 1 of the section titled [[Summary LFI|Summary of LFI data characteristics]].  
  
 +
=== Instrument scientific performance ===
  
: ''Back End Modules (BEM)''
+
==== ''Optical parameters''====
  
The BEMs are composed by four identical channels each made of Low Noise Amplifiers (LNA), RF Band Pass Filter, RF to DC diode detector and DC amplifiers.
+
The most accurate measurements of the LFI main beams were made with Jupiter, the most powerful unresolved (to Planck) celestial source in the LFI frequency range. Since the LFI feedhorns point to different positions on the sky, they detect the signal at different times.
 +
To map the beam, each sample in the selected timelines was projected in the (<i>u</i>, <i>v</i>) plane perpendicular to the nominal line-of-sight (LOS) of the telescope (and at 85&deg; to the satellite spin axis). The <i>u</i> and <i>v</i> coordinates are defined in terms of the usual spherical coordinates (&theta;, &phi;) :
  
In the 30 GHz BEM, each LNA consists of two cascaded MMIC amplifiers (HMC263 from Hittite).
+
: <math> u = \sin θ \, \cos φ, </math>
The Band Pass Filter is based on microstrip coupled line structure. Its design is a three order Chebyshev response band pass filter. Microstrip lines are made on Metclad substrate with thickness 0.254 mm and 2.17 dielectric relative constant.
+
: <math> v = \sin θ \, \sin φ. </math>
The detector is composed by a hybrid reactive/passive matching network, and a Schottky diode. A commercial Agilent beam-lead and zero-bias diode was selected (model HSCH-9161).
 
The detector diode is followed by a low noise OP27 operational amplifier that provides most of the DC amplification. A second stage is implemented using a OP200 operational amplifier to provide a balanced bipolar output.
 
  
In the 44 GHz BEM, each LNA consists of self designed MMIC amplifiers manufactured with the process ED02AH from OMMIC which employs a 0.2 μm gate length Pseudomorphic-High Mobility Transistor (P-HEMT) on GaAs.
+
To increase the signal-to-noise ratio, data were binned in an angular region of 2′ for the 70 GHz channels and 4′ for the 30 and 44 GHz channels. We recovered all beams down to −25 dB from the peak. An elliptical Gaussian was fit to each beam for both M and S radiometers. Differences between the M and S beams caused by optics and receiver non-idealities are inevitable at some level, but they appear to be well within the statistical uncertainties, and for the purposes of point-source extraction, the beams may be considered identical. For the details of the typical FWHM and ellipticity averaged over each frequency channel, refer to ({{PlanckPapers|planck2011-1-4 }}), ({{PlanckPapers|planck2014-a05||Planck-2015-A05}}) and {{PlanckPapers|planck2016-l02}}. Exhaustive details on all LFI beam parameters are presented in the LFI data processing section entitled [[Beams LFI|Beams]].
The topology chosen for the band-pass filter is a third order Chebyshef band pass filter made on a PTFE substrate (height=0.254mm, ε<math>_r</math>=2.17), based on microstrip coupled line structure.
 
The detector is composed by a hybrid reactive/passive matching network and a Schottky diode. A commercial Agilent beam-lead and zero-bias diode was selected (model HSCH-9161).  
 
The detector diode is followed by a low noise OP27 operational amplifier that provides most of the DC amplification. A second stage is implemented using a OP200 operational amplifier to provide a balanced bipolar output.
 
  
The 70 GHz BEM is constructed of machined aluminium with separate filter, amplifier/detector assemblies and an overall housing for other circuits and components.
+
==== ''Photometric Calibration''====
The FEM output signals are connected by waveguide from the Focal Plane Unit (FPU) assembly to the Back End Modules (BEM’s) housed adjacent to, the Data Acquisition Electronics (DAE) assembly. To maintain compatibility with the FEM’s, each BEM will accommodate, four receiver channels, from the four, waveguide outputs of each FEM. The BEM internal signal routes are not cross coupled and can be regarded as four identical parallel circuits.
 
Each BEM is constructed as two mirror halves. The two amplifier/detector assemblies each contain two amplifier/detector circuits. Each is supplied from one of a pair of printed circuit boards which also house two DC output amplifiers.
 
  
: ''4K Load''
+
Photometric calibration, i.e., conversion from voltage to antenna temperature, is performed for each radiometer after total power data have been cleaned of 1 Hz frequency spikes (see the LFI data processing section  [[TOI processing LFI#Spikes Removal|Spikes Removal]] page and {{PlanckPapers|planck2011-1-6 }}, {{PlanckPapers|planck2014-a03||Planck-2015-A03}}, {{PlanckPapers|planck2016-l02}} and differenced.
  
The purpose of the 4 K reference load is to provide the radiometer with a stable reference signal. Reducing the input offset (the radiometric temperature difference between the sky and the reference load) reduces the minimum achievable radiometer 1/f noise knee frequency for a given amplifier fluctuation spectrum. A reference load temperature that matches the sky temperature (approximately 2.7 K) would be ideal.
+
Here we report a brief overview of the photometric calibration; for details see the LFI data processing section [[TOI processing LFI#Photometric Calibration|Photometric Calibration]].
In the 4KRL design, the reference temperature is provided by the HFI outer radiation shield, at a temperature around 4K.
 
A subset of the 4KRL goal performance are reported in Table 5 below.
 
  
{| border="1" cellspacing="0" cellpadding="2" align="center"
+
Our calibrator is the dipole signal induced by spacecraft motion with respect to the CMB rest frame. The largest calibration uncertainty comes from the presence of the Galaxy and of the CMB anisotropies in the measured signal. We therefore use an iterative calibration procedure in which the dipole is fitted and subtracted, producing a sky map that is then removed from the original data to enhance the dipole signal for the next iteration. Typically, convergence is obtained after a few tens of iterations.  
|+ '''Table 4 4KLRL performance'''
 
|-
 
!scope="col"| Property
 
!scope="col"| HFI-mounted
 
|-
 
|width="200" | Physical Temperature 
 
|width="80" |< 5 K
 
|-
 
|width="200" | Mass
 
|width="80" |0.6 kg
 
|-
 
|width="200" |System reflectivity
 
|width="80" |< -20 dB
 
|-
 
|width="200" |System leakage
 
|width="80" |< -40 dB
 
|-
 
|width="200" |Required cooling power
 
|width="80" |< 1 mW
 
|-
 
|width="200" |Horn –load separation
 
|width="80" |> 1.5 mm
 
|-
 
|}
 
  
The 4K reference load unit is formed by single targets, one for each radiometer (two for each FEM). The horns used to couple to the 4 K reference load targets need to be relatively small because the targets themselves are small. An optimisation process produced a different horn design for each LFI band: their dimensions increase with reducing frequency. Due to the LFI Focal Plane design, where higher frequency radiometers (70 GHz) are placed around the HFI cryostat and the lower frequency radiometers (30 and 44 GHz) in a second row, the target mounting structure is separated in two parts, see Figure 5 below.
+
We calibrate in two major steps.
The upper one is located around the conical part of the HFI outer shield. Reference targets are mounted on a supporting structure, thermally and mechanically connected to the HFI outer shield. Each target faces a reference horn, two for each FEM. This ensemble is fixed to a support structure on the HFI 4K shield. Thermal link between the mounting structure and the HFI is obtained via fixation point only. Thermal washers are interposed to damp temperature fluctuations on targets induced by the HFI outer shield temperature oscillations.
+
First, we find the absolute calibration using the motion of the spacecraft around the sun (the 'orbital dipole'). This process uses only the temperature of the CMB and the very well known velocity of the spacecraft as inputs. From this calibration, and comparing with HFI, we can find the Planck estimate of the solar system velocity relative to the CMB (the 'solar dipole').
The lower part is fixed in the cylindrical part. It is made with the same target geometry of the upper part, and it is fixed on the HFI shield. The reference horns face the loads and are connected to the FEMs through WGs.
+
Second, We run another iterative calibration with this solar dipole as our signal. The larger signal gives us a better estimate of the time variation of our gain than that obtained from the orbital dipole.
Reference WGs and RH (Reference Horns) are either included in the FEM (70 GHz) or external to FEM (30 and 44
 
GHz).
 
  
[[File:4kload.jpg|thumb|center|400px|Figure 5. 4K reference load targets mounted on the HFI 4K shield]]
+
==== ''Noise Properties''====
  
Targets are formed of a back section, made of ECCOSORB CR117, which shows higher RF absorption but also high reflectivity. To reduce the target global reflectivity, a front section, assembled with an ECCOSORB specific cement to the back one, faces the radiometer Reference Horn. This last is casted from ECCOSORB CR110, whose RF reflectivity is lower than that of CR117. Target design is optimised to further reduce both reflectivity and leakage. Each target is metal backed and it is mounted in a metal enclosure.
+
The noise characteristics of the LFI data streams are closely reproduced by a simple (white + 1/<i>f</i> ) noise model,
  
=== ''REBA'' ===
+
<math> \label{eq:p1} P(f)= \sigma^2 \left[1+ \left(\frac{f}{f_{\rm knee}}\right)^\alpha \right] </math>,
  
The REBA is the electronic box in charge of processing the digitized scientific data and to manage the overall instrument. It is also in charge of the communication with the spacecraft.
+
where <i>P</i>(<i>f</i>) is the power spectrum and &alpha;&asymp;−1.
There are two REBA boxes, one nominal and one redundant. The redundancy concept is cold, which means that both boxes are never ON at the same time; the operation of each unit shall be managed by the spacecraft switching-on the corresponding unit. The REBA_ASW is the same in each REBA box.
+
In this model, noise properties are characterized by three parameters, the white noise limit &sigma;, the knee frequency <i>f</i><sub>knee</sub>, and the exponent of the 1/<i>f</i> component &alpha;, also referred to as the slope. Here we give noise performance estimates based on one year of operations; details of the analysis are given in LFI data processing section ([[TOI processing LFI#Noise|Noise estimation]]) and in  
 +
({{PlanckPapers|planck2011-1-6}}, {{PlanckPapers|planck2011-1-4 }}, {{PlanckPapers|planck2014-a03||Planck-2015-A03}} and {{PlanckPapers|planck2016-l02}}.
  
Each REBA consists of the following subunits:
+
Noise properties have been calculated following two different and complementary approaches: 1) fitting the equation above to time-ordered data for each radiometer; and 2) building normalized noise maps by differencing data from the first half of each pointing period with data from the second half of that pointing period to remove the sky signal (“jackknife” data sets).
  
− the Power Supply Unit (PSU) which feeds the REBA unit. It consists of a DC/DC converter that converts the primary power received from the spacecraft PDU to the secondary regulated voltages required only by the REBA and provides galvanic isolation towards the spacecraft side of the interface. The PSU DC/DC converter also receives the On-Board Clock (OBC) from the CDMS that is used to increment the internal On Board Time register. There is no software interface with the REBA_ASW.
+
Typical uncertainties are 0.5% for the white noise, between 5 and 10% for the slope, and between 10 and 20% for the knee frequency.
  
− the Data Acquisition Unit (DAU) performs the analogue to digital conversion of the analogue housekeeping data of the REBA itself (temperatures and voltages). The REBA_ASW collects from the DAU the HK data.
+
==== ''White Noise Sensitivity''====
  
− the Signal Processing Unit (SPU) is a computing subunit and implements part of the REBA_ASW, the SPU_ASW. It receives the science data from the DAE through a IEEE 1355 link implemented in a SMCS chip. A second IEEE 1355 link is used to control by link the remote DAE SMCS chip. The third IEEE 1355 link communicates with the DPU. The SPU is in charge of the reduction and compression of the science data.
+
Details of the white noise sensitivity can be found in ({{PlanckPapers|planck2011-1-4}}).  
The SPU hardware of LFI consists of one CPU, which main characteristics are TSC21020E processor based architecture running up to 18 MHz; PROM Memory for Bootstrap Program Storage (192 KB); SRAM Memory for Program Storage (3 MB), protected by EDAC; SRAM Memory for Data Storage (2 MB), protected by EDAC; Expansion Bus of 32 bit mapped at the Data Memory Bus address space. An SRAM Memory Data Expansion Module (2 MB), protected by EDAC, are connected to the CPU; three IEEE-1355 DS High Speed Serial Communication links, implemented by the SMCS332 chip and including a (32 KB) Dual Port Ram buffer; programmable Watchdog Timer.
 
Two of the IEEE-1355 links are used by the SPU_ASW to communicate with the DAE (scientific data) and to control by link the DAE SMCS chip; the third one communicates with the DPU (commands and TM).
 
A “Data Ready” electrical signal is connected between DAE and the SPU, this signal produces an interruption in the SPU when the DAE is ready to transfer data.
 
The SPU_ASW is stored in the EEPROM located in the DPU board (see later description). It shall be transferred to the SPU by the DPU_ASW.
 
  
the Digital Processing Unit (DPU) is a computing subunit and implements part of the REBA_ASW, the DPU_ASW. It contains another SMCS chip with 3 IEEE 1355 links that communicate with the SPU, and the DAE. A MIL-STD 1553B link is used to communicate with the CDMS. The DPU is in charge of the control and monitoring of the instrument as well as the communication with the spacecraft (CDMS).
+
Table 2 summarizes the sensitivity numbers calculated during the first year of operations using methods and procedures described in detail in ({{PlanckPapers|planck2011-1-6}}, {{PlanckPapers|planck2014-a03||Planck-2015-A03}} and {{PlanckPapers|planck2016-l02}}, compared with scientific requirements. The measured sensitivity is in very good agreement with pre-launch expectations. While the white noise moderately exceeds the design specification, this performance is fully in line with the LFI science objectives.
The DPU hardware of LFI consists of one CPU, whose main characteristics are a TSC21020E processor based architecture running up to 18 MHz; PROM Memory for Bootstrap Program Storage (192KB); SRAM Memory for Program Storage (3 MB), protected by EDAC;  SRAM Memory for Data Storage (2 MB), protected by EDAC; three IEEE-1355 DS High Speed Serial Communication links, implemented by the SMCS332; chip and including an 8Kx32 bits (32 KB) Dual Port Ram buffer; one MIL-STD-1553B Data Bus Interface (redundant A+B) including a (16KB) Dual Port Ram Buffer; Programmable Watchdog Timer;  One S/C OBT interface (131.072 KHz clock signal); one Back End Unit Timing interface (1 Hz signal); an EEPROM (1.5 MB), protected by EDAC, for on-board storage of the SPU_ASW and DPU_ASW.
 
One IEEE 1355 link is used by the DPU_ASW to communicate with the DAE to control by link the SMCS chip of DAE. The second one is used to communicate with the DAE to transfer commands and HK and the third one is used by DPU_ASW to communicate with the SPU (commands and TM). Two Reset electrical lines are provided by the DPU to reset each of the two SMCS chips of DAE. The DPU_ASW is stored in the EEPROM.
 
  
=== ''Instrument On-board Software'' ===
+
{| border="1" cellspacing="0" cellpadding="2" align="center"
 +
|+ '''<small>Table 2. White noise sensitivity of the LFI frequency channels compared with requirements.</small>'''
 +
|-
 +
!scope="col"| Channel
 +
!scope="col"| Measurement [&mu;K<sub>CMB</sub> s<sup>1/2</sup>]
 +
!scope="col"| Requirement [&mu;K<sub>CMB</sub> s<sup>1/2</sup>]
 +
|-
 +
|width="120" | 70 GHz
 +
|width="100" | 152.0
 +
|width="100" | 119
 +
|-
 +
|width="120" | 44 GHz
 +
|width="100" | 174.2
 +
|width="100" | 119
 +
|-
 +
|width="120" | 30 GHz
 +
|width="100" | 148.1
 +
|width="100" | 119
 +
|-
 +
|}
  
=== ''Instrument Operations'' ===
+
=== Instrument technical performance ===
  
  
=== LFI Naming Convention ===
+
==== ''Spectral response''====
  
The naming of all the LFI elements is described in the previous sections but here is summarized again for clarity.  
+
The in-band receiver response was thoroughly modelled and measured for all the LFI detectors during ground tests. The complete set of bandpass curves was published in (Zonca et al. 2009{{BibCite|zonca2009}}, {{PlanckPapers|planck2014-a06||Planck-2015-A06}}) where all the details of the LFI radiometer's spectral response are given. From each curve we have derived the effective centre frequency according to:
  
The 11 RCAs are labelled by a numbers from 18 to 28 as outlined in Figure 1, right panel. In each RCA, the two perpendicular linear polarisation components are labelled as ''M'' or ''S'' according to the arm of the OMT they are connected to (''Main'' or ''Side'', see lower-left inset of Figure 3).
+
<math> \label{eq:spectr}
 +
\nu_0 = \frac{ \int_{\nu_{\rm min}}^{\nu_{\rm max}} \nu g(\nu) \; d \nu}
 +
{\int_{\nu_{\rm min}}^{\nu_{\rm max}} g(\nu)\; d\nu }
 +
</math>
  
Each front-end amplifier (see upper-left inset of Figure 3) is labelled with codes ''1'', ''2'' so that the four outputs of the LNAs can be named with the sequence: ''M1'', ''M2'' (radiometer ''M'') and ''S1'', ''S2'' (radiometer ''S'').
+
where &Delta;&nu; = &nu;<sub>max</sub> − &nu;<sub>min</sub> is the receiver bandwidth and <i>g</i>(&nu;) is the bandpass response.
 +
Details about colour corrections, <i>C</i>(&alpha;), needed to derive the brightness temperature of a source with a power-law spectral index &alpha;, are provided in
  
Each radiometer has two output diodes which are labelled with binary codes ''00'', ''01'' (radiometer ''M'') and ''10'', ''11'' (radiometer ''S''), so that the four outputs of each radiometric chain can be named with the sequence: ''M-00'', ''M-01'', ''S-10'', ''S-11''.
+
Some details are also given in the corresponding section of the [[LFIAppendix#Spectral Response|Annexes]].
  
  
 +
===== ''Bandpass estimation''=====
  
== LFI Ground Tests ==
+
As detailed in (Zonca et al. 2009{{BibCite|zonca2009}}, {{PlanckPapers|planck2014-a06||Planck-2015-A06}}), our most accurate method to measure the LFI bandpasses is based on measurements of individual components integrated into the LFI Advanced RF Model (LARFM) to yield a synthesized radiometer bandpass.
 +
The LARFM is a software tool based on the open-source Quasi Universal Circuit Simulator (QUCS). The measured frequency responses of the various subsystems (feed-OMT, FEM, BEM) are considered as lumped S-parameter components.
 +
Measurements of single components are obtained with standard methods and provide highly reliable results, with precision of order 0.1-0.2 dB over the entire band.
 +
Waveguides are simulated with an analytical model, in order to reproduce the effect of their temperature gradient and the effect of standing waves caused by impedance mismatch at the interfaces between the FEM and BEM. This is because the 1.8-m long waveguides were not measured at unit level in cryogenic conditions. The model provides accurate agreement with the measured waveguide response in the conditions of the test measurements (300 K).
 +
The composite bandpasses are estimated to have a precision of about 1.5 to 2 dB.
  
During its development, the LFI flight model was calibrated and tested at various integration levels from sub-systems [R. J. Davis, A. Wilkinson, R. D. Davies et al. Design, development and verification of the 30 and 44 GHz front-end modules for the Planck Low Frequency Instrument. Journal of Instrumentation, 4:2002, 2009] to individual integrated receivers [F. Villa, L. Terenzi, M. Sandri et al. Planck pre-launch status: Calibration of the Low Frequency Instrument flight model radiometers. A&A, 520:A6, 2010] and the whole receiver array [A. Mennella, M. Bersanelli, R. C. Butler et al. Planck pre-launch status: Low Frequency Instrument calibration and expected scientific performance. A&A, 520:A5, 2010]. In every campaign we performed tests according to the following classification:
+
Some details are also given in the corresponding section of the [[LFIAppendix#Bandpass Estimation|Annexes]].
  
* ''Functionality tests'', performed to verify the instrument functionality.
+
==== ''Stability''====
* ''Tuning tests'', to tune radiometer parameters (biases, DC electronics gain and offset, digital quantisation and compression) for optimal performance in flight-like thermal conditions.
 
* ''Basic calibration and noise performance tests'', to characterise instrument performance (photometric calibration, isolation, linearity, noise and stability) in tuned conditions.
 
* ''Susceptibility tests'', to characterise instrument susceptibility to thermal and electrical variations.
 
  
Where possible, the same tests were repeated in several test campaigns, in order to ensure enough redundancy and confidence in the instrument behaviour repeatability. A critical comparison, that is central to the subject of this work, is the one between the results of on-ground and in-flight test campaigns. A matrix showing the instrument parameters measured in the various test campaigns is provided in Table 1 of [A. Mennella, M. Bersanelli, R. C. Butler et al. Planck pre-launch status: Low Frequency Instrument calibration and expected scientific performance. A&A, 520:A5, 2010].
+
Thanks to its differential scheme, the LFI is insensitive to many effects caused by 1/<i>f</i> noise, thermal fluctuations, or electrical instabilities.
 +
As detailed in ({{PlanckPapers|planck2011-1-4}}), one effect detected during the first survey was the daily temperature fluctuation in the back-end unit induced by the downlink transponder, which was powered on each day for downlinks during the first 258 days of the mission. As expected, the effect is highly correlated between the sky and reference load signals. In the difference, the variation is reduced by a factor around (1 − <i>r</i>), where <i>r</i> is the gain modulation factor defined above (see <i>r</i> definition in [[LFI design, qualification, and performance#Radiometer Chain Assembly (RCA)|RCA]] section).
  
The ground test campaign was developed in three main phases: cryogenic tests on the individual RCAs, cryogenic tests on the integrated receiver array (the so-called radiometer array assembly, RAA) and system-level tests after the integration of the LFI and HFI instruments onto the satellite. The first two phases were carried out at the Thales Alenia Space - Italia laboratories located in Vimodrone (Milano, Italy) (note that receiver tests on 70 GHz RCAs were carried out in Finlad, at Yilinen laboratories), system level tests (SLT) were conducted in a dedicated cryofacility at the Centre Spatiale de Liege (CSL) located in Liege (Belgium).  
+
A particular class of signal fluctuations occasionally observed during operations was due to electrical instabilities that appeared as abrupt increases in the measured drain current of the front-end amplifiers, with a relaxation time variable from few seconds to some hundreds of seconds. Typically, these events caused a simultaneous change in the sky and reference load signals. Because they are essentially common-mode, their residual on the differenced data is negligible, and the data are suitable for science production. In a few cases the residual fluctuation in the differential output was large enough (a few millikelvin in calibrated antenna temperature units) to be flagged, and the data were not used. The total amount of discarded data for all LFI channels until Operational Day 389 was about 2000s per detector, or 0.008%.
  
In Table 1 below we list the temperature of the main cold thermal stages during ground tests compared to in-flight nominal values. These values show that system-level tests were conducted in conditions that were as much as possible flight-representative, while results obtained during RCA and RAA tests needed to be extrapolated to flight conditions to allow comparison. Details about the RCA test campaign are discussed in [F. Villa, L. Terenzi, M. Sandri et al. Planck pre-launch status: Calibration of the Low Frequency Instrument flight model radiometers. A&A, 520:A6, 2010] while the RAA tests and the extrapolation methods are presented in [A. Mennella, M. Bersanelli, R. C. Butler et al. Planck pre-launch status: Low Frequency Instrument calibration and expected scientific performance. A&A, 520:A5, 2010].
+
A further peculiar effect appeared in the 44 GHz detectors, where single isolated samples, either on the sky or the reference voltage output, were clear outliers compared with the rest. Over a reference period of four months, 15 occurrences of single-sample spikes (out of 24 total anomaly events) were discarded, an insignificant loss of data.
   
 
[[File:groundtest.jpg|thumb|center|400px|Table 1. Temperatures of the main cold stages during the various ground test campaigns compared to in-flight nominal values.]]
 
  
Cryogenic system-level tests were split into three parts:
+
==== ''Thermal susceptibility''====
* ''Thermal balance'', to validate the overall thermal mathematical model. The ground vacuum test equipment simulated the space environment.
 
* ''System cryogenic test'', to check and optimise the satellite and instrument performance working at nominal temperatures.
 
* ''Thermal cycling test'', to check the reliability of all electronics equipment and instruments to temperature variations (note that to allow the thermal control of the Service Vehicle Module (SVM), the warm part of the spacecraft was surrounded by a cryogenic shroud at a temperature lower than 100 K. The temperature of the shrouds was adjustable in order to allow the thermal cycling of the SVM during the tests. The temperature ranged between the two extreme values of <math>20^\circ</math>C to <math>100^\circ </math>C).
 
  
During the various test campaigns the instrument was switched off and moved several times in a time period of about three years. A series of functional tests were always repeated at each location and also in flight, in order to verify the instrument functionality and the response repeatability. No failures or major problems have been identified due to transport and integration procedures.
+
As mentioned in the section [[LFI design, qualification, and performance#LFI In-flight Calibration|LFI In-flight Calibration]] above, and detailed in (Gregorio et al. 2013{{BibCite|gregorio2013}}), during the CPV campaign, susceptibility tests were performed in order to characterize the LFI instrument susceptibility to thermal and electrical variations.
  
 +
The effect of temperature fluctuations on the LFI radiometers originates in the Planck cold end interface of the hydrogen sorption cooler to the instrument focal plane. The temperature is actively controlled through a dedicated stage, the Thermal Stabilization Assembly (TSA), providing a first reduction of the effect. The thermal mass of the focal plane strongly contributes to reduce residual fluctuations.
 +
The physical temperature fluctuations propagated at the front end modules cause a correlated fluctuation in the radiometer signal, degrading the quality of scientific data. The accurate characterization of this effect is crucial for attempts to remove it from raw data by exploiting the housekeeping information on thermal sensors.
  
== LFI In-flight Calibration ==
+
The propagation of the temperature oscillations through the focal plane and the instrument response to thermal changes were characterized through two main tests:
 +
* the thermal dynamic response, aimed at measuring the dynamic thermal behaviour of the LFI Focal Plane;
 +
* the thermal susceptibility of the radiometers.
  
The Planck cryo-chain [Planck Collaboration. Planck early results. II. The thermal performance of Planck. A&A, 536:A2,
+
Further details are also given in the corresponding section of the [[LFIAppendix#Thermal Susceptibility|Annexes]].
2011] is composed by three coolers: a 20 K Hydrogen sorption cooler, which is passively pre-cooled by the <math>3^{\rm rd}</math> V-groove radiator, cools the LFI focal plane and pre-cools the HFI 4K cooler; a Stirling 4K cooler that cools the HFI box and feed-horns and provides a 4K blackbody reference signal to the LFI receivers; a 0.1 K dilution cooler, which is pre-cooled by the 4K cooler and cools the HFI bolometer filters to <math>\sim 1.6</math>K and the HFI bolometer detectors to ~0.1 K. The cooldown of the HFI 4K stage, in particular, was key during CPV for the LFI as it provided a variable input signal that was exploited during bias tuning.
 
  
The LFI CPV started on June, <math>4^{\rm th}</math> 2009 and lasted until August, <math>12^{\rm th}</math> when Planck started scanning the sky in nominal mode. At the onset of CPV, the active cooling started when the radiating surfaces on the payload module reached their working temperatures (~50 K on the <math>3^{\rm rd}</math> V-groove, and ~40 K on the reflectors) by passive cooling. This was achieved during the transfer phase. Nominal temperatures were achieved on July, <math>3^{\rm rd}</math> 2009, when the dilution cooler temperature reached 0.1 K [Planck Collaboration. Planck early results. II. The thermal performance of Planck. A&A, 536:A2, 2011; P. Meinhold, R. Leonardi, B. Aja et al. Noise properties of the Planck-LFI receivers. Journal of Instrumentation, 4:2009, 2009] (see Figure 5 below).
+
==== Instrument budgets ====
  
The CPV was carried out in four phases (see Table 2 below for a summary of the overall CPV test campaign):
+
LFI power, mass and telemetry budgets are given in the corresponding section of the [[LFIAppendix#Instrument Budgets|Annexes]].
* LFI switch on and basic functionality verification,
 
* Tuning of front-end biases and back-end electronics,
 
* Preliminary calibration tests,
 
* Thermal tests.  
 
  
[[File:flighttest.jpg|thumb|center|400px|Table 2.]]
+
== <span id="LFISystematics">Systematic Effects </span>==
  
== LFI Performance ==
+
The LFI design was driven by the need to suppress systematic effects well below instrument white noise. The differential receiver scheme, with reference loads cooled to 4 K, greatly minimizes the effect of 1/<i>f</i> noise and common-mode fluctuations, such as thermal perturbations in the 20-K LFI focal plane. The use of a gain modulation factor (see <i>r</i> definition in the [[LFI design, qualification, and performance#Radiometer Chain Assembly (RCA)|RCA]] section above) largely compensates for spurious contributions from input offsets. Furthermore, diode averaging (see <i>V</i><sup>rad</sup><sub>out</sub> definition in the [[LFI design, qualification, and performance#Radiometer Chain Assembly (RCA)|RCA]] section above) allows us to cancel second-order correlations, such as those originating from phase switch imbalances.
  
 +
We have developed an error budget for systematic effects ({{PlanckPapers|bersanelli2010}}, {{PlanckPapers|planck2011-1-4}}{{PlanckPapers|planck2013-p02a}} {{PlanckPapers|planck2014-a04||Planck-2015-A04}} and {{PlanckPapers|planck2016-l02}}) as a reference for both instrument design and data analysis. Our goal is to ensure that each systematic effect is rejected to the specified level, either by design or by robust removal in software. At this stage, the following effects are relevant:
  
=== Instrument Scientific Performance ===
+
[[File:Selection_258.png|500px]]
  
 +
For each of these effects we used flight data and information from ground tests to build timelines, maps, and angular power spectra that represent our best knowledge of their impact on the scientific analysis.
 +
The details of the systematic effect analysis are given in the section entitled LFI data processing section ([[LFI systematic effect uncertainties|Systematic Effects uncertainties]]).
  
=== Instrument Technical Performance and Budgets'' ===
+
== <span id="SCS">The Sorption Cooler</span> ==
 +
The Planck H2 Sorption Cooler is the first stage of the active cryogenic chain. Its task is to maintain the LFI down to the operating temperature while providing a pre-cooling stage for the HFI refrigerators. The system performs a simple thermodynamic cycle based on hydrogen compression, gas pre-cooling by three passive radiators, further cooling due to the heat recovery by the cold low pressure gas stream, expansion through a J-T valve, and evaporation at the cold stage ({{PlanckPapers|planck2011-1-3}}). A schematic of the Planck Sorption Cooler System (SCS) is shown in Fig. 5. The engine of the cryocooler is the compressor.  It serves two main functions: to produce the high-pressure hydrogen gas flow; and to maintain a stable gas recovery rate, which keeps the return pressure, hence the liquid temperature, constant. The high pressure gas flowing to the cold end is pre-cooled by exchanging heat with the three passive stages at the V-Grooves and with the evaporated cold gas returning back to the compressor. The gas then expands at the cold end through a J-T valve, producing approximately 1 Watt of cooling power at a temperature of <20K: most of this heat lift is used at the LVHX2 (Liquid Vapour Heat eXchanger 2, see Fig. 5) interface to absorb the LFI heat load at a temperature around 20K. The remaining heat lift is used at the LVHX1 as a pre-cooling stage, at a temperature lower than 19 K, for the two HFI refrigerators. The cooler and its performance are described in detail in (Morgante et al. 2009{{BibCite|morgante2009}}).
  
 +
[[File:SCS_CAD_Figure.jpg|thumb|center|480px|'''Figure 5. Planck Sorption Cooler CAD view with labels of the main sub-systems.''']]
  
== LFI Systematics Effects ==
+
Both LVHXs provide a temperature of around 18 K, with fluctuations driven by the cooler instabilities (compressor element variations, cycling, two-phase flow dynamics, etc.). Stabilization of the HFI interface (LVHX1) temperature is not necessary, since thermal control of the subsequent colder stages is more efficient and very effective.  To reduce cold end fluctuations directly transmitted to the LFI radiometers, a copper block, designated as the Temperature Stabilization Assembly (TSA), is inserted, as an intermediate stage, between LVHX2 and the LFI Focal Plane Unit (FPU). The TSA comprises a temperature sensor and a heater, controlled by a PID-type feedback loop, working in combination with the passive thermal inertia. The set-point temperature of the TSA is an adjustable parameter of the sorption cooler system, chosen to provide dynamic range for control during mission: as the compressor elements age, the return gas pressure and thus the temperature of LVHX2 rise slowly. To keep the LFI temperature reference stable during operations the set point must be periodically adjusted to maintain the level of oscillations within the required range (Fig. 6).
  
 +
[[File:TSA_LVHX2_mission.jpg|thumb|center|480px|'''Figure 6. LVHX2 (black, HK parameter name SD029540) and TSA (red, HK parameter name SD030540) temperature profile from 125 to 563 days after launch.''']]
  
 +
Two sorption cooler units were integrated on-board the Planck spacecraft. Both units were used to cover the mission lifetime, with one cooler operated for the first two sky surveys. At the end of its lifetime a switchover operation, performed on August 11th 2010 (455 days after launch), activated the second unit that ran since then. The LVHX2 temperature change due to the new cooler start, and the subsequent re-adjustment of the LFI temperature stabilization stage, is clear in Fig. 6.
  
 +
<!--
 
== Acronyms ==
 
== Acronyms ==
  
== The Bibliography ==
+
: ADC Analog-to-Digital Converter
 +
: ADU Analog-to-Digital Unit
 +
: BEM Back End Module
 +
: BEU Back End Unit
 +
: CCE Central Check-out Equipment
 +
: CDMS Command and Data  Management Subsystem
 +
: CDMU Central Data Management Unit
 +
: CoG Centre of Gravity
 +
: CPV Calibration and Performance Verification
 +
: CSL Centre Spatiale de Liege
 +
: DAE Data Acquisition Electronics
 +
: DC Direct Current
 +
: DPC Data Processing Centre
 +
: DPU Digital Processing Unit
 +
: EMC Electro-Magnetic Compatibility
 +
: EMI Electro-Magnetic Interference
 +
: FEM Front End Module
 +
: FEU Front End Unit
 +
: FH Feed Horn
 +
: FOV Field Of View
 +
: FPU Focal Plane Unit
 +
: HK House Keeping
 +
: ILT Instrument Level Test
 +
: IST Integrated System Test
 +
: JFET Junction Field Effect Transistor
 +
: LEOP Launch and Early Orbit Phase
 +
: MLI Multilayer Insulation
 +
: MoI Moment of Inertia
 +
: MOS Margin Of Safety
 +
: OMT Ortho Module Transducer
 +
: PCS Power Control Subsystem
 +
: PSF Point Spread Function
 +
: RAA Radiometer Array Assembly
 +
: RAM Random Access Memory
 +
: RCA Radiometer Chain Assembly
 +
: REBA Radiometer Electronics Box Assembly
 +
: S/C Spacecraft
 +
: SCC Sorption Cooler Compressor assembly
 +
: SCCE Sorption Cooler Cold End
 +
: SCE Sorption Cooler Electronics
 +
: SCOS Spacecraft Control and Operations System
 +
: SCP Sorption Cooler Piping
 +
: SCS Sorption Cooler Subsystem
 +
: SLT System Level Test
 +
: SPU Signal Processing Unit
 +
: SS Stainless Steel
 +
: SVM Service Module
 +
: TCS Thermal Control System
 +
: TM Telemetry
 +
: TSA Thermal Stabilization Assembly
 +
: TTC Telemetry, Tracking and Command
 +
: VSWR Voltage Standing Wave Ratio
 +
: WG Waveguide
 +
 
 +
== Glossary ==
 +
 
 +
: Feed Horns xxx
 +
: REBA Radiometer Electronics Box Assembly
 +
-->
 +
 
 +
== References ==
 +
 
 +
<References />
  
  
<biblio force=false>
+
[[Category:LFI design, qualification and performance|001]]
#[[References]]
 
</biblio>
 

Latest revision as of 15:28, 4 July 2018

Instrument description[edit]

The Low-Frequency Instrument (see Fig. 1 in the LFI overview) consists of a 20 K focal plane unit hosting the corrugated feed horns, the orthomode transducers (OMTs) and the receiver front-end modules (FEMs). Forty-four composite waveguides [1] are interfaced with three conical thermal shields and connect the front-end modules to the warm (roughly 300 K) back-end unit (BEU) containing a further radio frequency amplification stage, detector diodes, and all the electronics for data acquisition and bias supply.

Best LFI noise performance is obtained with receivers based on InP High Electron Mobility Transistor (HEMT) low noise amplifiers (LNAs) for minimal power dissipation and best performance. To further minimize power consumption in the focal plane, the radiometers are split into two sub-assemblies connected by waveguides, one located at the telescope focal area, the other on the 300 K portion of the Planck satellite. These design features allow the entire front-end LNA dissipation to be <0.55 W, which enables the active cooling of the focal assembly. This is achieved with a vibration-less hydrogen sorption cooler, which also provides 18 K pre-cooling to the HFI helium J-T cooler. Two sorption cooler units are included in the flight hardware.

As shown schematically in Fig. 1 below, the LFI consists of the following subsystems:

  • Radiometer Array Assembly (RAA);
  • Sorption Cooler Subsystem (SCS);
  • Radiometer Electronics Box Assembly (REBA).

The RAA includes the Front End Unit (FEU) and the Back End Unit (BEU), connected via waveguides. The FEU is located at the focus of the telescope, as one component of the joint LFI/HFI focal assembly (see sections below). The BEU is mounted on the top of the Planck service module (SVM). The REBA (Radiometer Electronics Box Assembly) and the warm parts of the Sorption Cooler System (SCS) are located on one of the lateral panels of the SVM. The FEU and the Sorption Cooler Compressor (SCC) are connected by concentric stainless steel tubes. The smaller tube carries hydrogen at approximately 60 atmospheres from the cooler compressors to the FEU, while the larger tube returns the hydrogen at about 0.3 atmospheres. These units are described in following sections and in the Annexes, while the SCS is described in details in the Sorption Cooler section. All LFI units are linked together by the LFI harness, which also connects to the spacecraft interface.

Figure 1. Block Diagram of LFI.

Radiometer Array Assembly (RAA)[edit]

The Radiometer Array Assembly (RAA) consists of two main units (the front end unit, FPU and the Back End Unit, BEU), connected by a set of waveguides. The Focal Plane Unit (FPU) is the heart of the LFI instrument; it contains the feed array and associated orthomode transducers (OMTs) and FEMs, all cooled to 20 K by the sorption cooler. The FPU comprises a set of 11 modules, which are mounted on a mechanical support which meets the thermo-mechanical requirements of the instrument and adds thermal inertia. The BEU comprises the radiometer Back End Modules (BEM) and the Data Acquisition Electronics (DAE), which are connected by an internal harness. The HFI Unit is located inside the LFI FPU and supported by the LFI structure. The LFI structure gives the mechanical and thermal interface to the HFI unit with the proper stiffness and thermal de-coupling. The LFI structure also guarantees the proper alignment of the HFI detector with the telescope focal plane.

The timescale of the stability of the receiverS is driven by the 1 rpm rotation speed of the spacecraft, which requires a very low 1/f-noise or gain variation of the low noise amplifiers and other components. The LFI uses a pseudo-correlation receiver concept (Fig. 2 below). This radiometer concept is chosen to maximize the stability of the instrument by reducing the effect of non-white noise generated in the radiometer itself. In this scheme, the difference between the inputs to each of the chains (the signal from the telescope and that from a reference blackbody, respectively) is continuously being observed. To remove the effect of instability in the back-end amplifiers and detector diodes, it is necessary to switch the signal detected at the diodes at a high rate. The signals from the sky and from a reference load are combined by a hybrid coupler, amplified in two independent amplifier chains, and separated out by another hybrid. The sky and the reference load power can then be measured and differenced. Since the reference signal has been subject to the same gain variations in the two amplifier chains as the sky signal, the true sky power can be recovered. The differencing receiver greatly improves the stability if the two input signals are almost equal, at a cost of a factor of √2 in sensitivity, compared to a perfectly stable total-power radiometer with the same noise temperature and bandwidth. This radiometer concept is also capable of greatly reducing the knee frequency. A single Radiometer Chain Assembly (RCA; see Fig. 2) consists of each functional unit from the feed horn to the BEM. The RAA therefore includes a set of 11 RCAs and the Data Acquisition Electronics (see also Fig. 1 above), all mounted on a suitable mechanical structure. Although there are differences in the details of the radiometer chains at different frequencies, their overall configuration is similar, and a general description of their design is provided in this section. Planck LFI has 11 Radiometer Chain Assemblies (RCAs). Each RCA is constituted by a feed horn and FEM in the FEU (at 20 K), BEM (at 300 K) in the BEU, and four waveguides that connect each FEM-BEM couple. The frequency distribution of the RCA is the following:

  • 2 RCAs at 30 GHz;
  • 3 RCAs at 44 GHz;
  • 6 RCAs at 70 GHz.
Figure 2. A complete RCA from feed-horn to analogue voltage output. The insets show the OMT, the details of the 20 K pseudo-correlator and of the back-end radio-frequency amplification, low-pass filtering, detection, and DC amplification.

Radiometer Chain Assembly (RCA)[edit]

Every RCA consists of two radiometers, each feeding two diode detectors (see Fig. 2 above), for a total of 44 detectors. The 11 RCAs are labelled by a number from 18 to 28, as outlined in Fig. 1 of the LFI overview, right panel.

Fig. 2 provides a more detailed description of each radiometric receiver. In each RCA, the two perpendicular linear polarization components split by the OMT propagate through two independent pseudo-correlation differential radiometers, labelled as M or S, depending on the arm of the OMT they are connected to (Main or Side, see lower-left inset of Fig. 2).

In each radiometer the sky signal coming from the OMT output is continuously compared with a stable 4 K blackbody reference load mounted on the external shield of the HFI 4-K box (Valenziano et al. 2009[2]). After being summed by a first hybrid coupler, the two signals are amplified by approximately 30 dB (see upper-left inset of Fig. 2). The amplifiers were selected for best operation at low drain voltages and for gain and phase match between paired radiometer legs, which is crucial for good balance. Each amplifier is labelled with codes 1 or 2 so that the four outputs of the LNAs can be named with the sequence: M1, M2 (radiometer M) and S1, S2 (radiometer S). Tight mass and power constraints called for a simple design of the Data Acquisition Electronics (DAE) box so that power bias lines were divided into five common-grounded power groups with no bias voltage readouts; only the total drain current flowing through the front-end amplifiers is measured and is available to the house-keeping telemetry (this design has important implications for front-end bias tuning, which depends critically on the satellite electrical and thermal configuration and was repeated at all integration stages during on-ground and in-flight satellite tests). A phase shift (or phase switch) alternating between 0° and 180° at the frequency of 4096 Hz is applied in one of the two amplification chains and then a second hybrid coupler restores the sky and reference load components, which are further amplified and detected in the warm BEU, with a voltage output ranging from -2.5 V to +2.5 V.

Each radiometer has two output diodes which are labelled with binary codes 00, 01 (radiometer M) and 10, 11 (radiometer S), so that the four outputs of each radiometric chain can be named with the sequence M-00, M-01, S-10, S-11.

After detection, an analogue circuit in the DAE box removes a programmable offset in order to obtain a nearly null DC output voltage and a programmable gain is applied to increase the signal dynamics and optimally exploit the ADC input range. After the ADC, data are digitally down-sampled, re-quantized and compressed in the REBA according to a scheme described in (Herreros et al. 2009[3], Maris et al. 2009[4]), before being combined in telemetry packets. On the ground, telemetry packets are converted to time ordered data for both the sky and reference load, after calibrating the ADU (Analog-to-Digital Unit) samples into volts. The calibration takes into account the applied offset and gain factors.

To first order, the mean differential power output for each of the four receiver diodes can be written as follows (Seiffert et al. 2002[5], Mennella et al. 2003[6], Planck-PreLaunch-IV[7]):

[math] \label{eq:power} P_{\rm out}^{\rm diode} = a\, G_{\rm tot}\,k\,\beta \left[ T_{\rm sky} + T_{\rm noise} - r\left( T_{\rm ref} + T_{\rm noise}\right) \right] [/math]

where Gtot is the total gain, k is the Boltzmann constant, β the receiver bandwidth, and a is the diode constant. Tsky and Tref are the average sky and reference load antenna temperatures at the inputs of the first hybrid and Tnoise is the receiver noise temperature.

The gain modulation factor (Mennella et al. 2003[6], Planck-Early-V[8]), r, is defined by

[math] \label{eq:erre1} r = \frac{T_{\rm sky} + T_{\rm noise}}{T_{\rm ref} + T_{\rm noise}} , [/math]

and is used to balance (in software) the temperature offset between the sky and reference load signals and minimize the residual 1/f noise in the differential data stream. This parameter is calculated from the average uncalibrated total power data using the relationship

[math] \label{eq:erre2} r = \langle V_{\rm sky} \rangle/ \langle V_{\rm ref}\rangle, [/math]

where <Vsky> and <Vref> are the average sky and reference voltages calculated in a defined time range. The white noise spectral density at the output of each diode is essentially independent of the reference-load absolute temperature and is given by

[math] \label{eq:dt1} \Delta T_0^{\rm diode} = \frac{2\,(T_{\rm sky}+T_{\rm noise})}{\sqrt{\beta}}. [/math]

If the front-end components are not perfectly balanced, then the separation of the sky and reference load signals after the second hybrid is not perfect and the outputs are mixed. First-order deviations in white noise sensitivity from the ideal behaviour are caused mainly by noise temperature and phase-switch amplitude mismatches. Following the notation used in (Seiffert et al. 2002 [5]), we define εTn, as the imbalance in front end noise temperature, and εA1 and εA2, the imbalance in signal attenuation in the two states of the phase switch. Equation above for the two diodes of a slightly imbalanced radiometer then becomes

[math] \label{eq:dt2} (\Delta T^{\rm diode} )^2 ≈ (\Delta T_0^{\rm diode})^2 ( 1± \frac{\epsilon_{A1}- \epsilon_{A2}}{2} + \alpha \epsilon_{T{\rm n}}), [/math]

which is identical for the two diodes apart from the sign of the term εA1 − εA2, representing the phase switch amplitude imbalance. This indicates that the isolation loss caused by this imbalance generates an anti-correlation between the white noise levels of the single-diode data streams. For this reason, the LFI scientific data streams are obtained by averaging the voltage outputs from the two diodes in each radiometer:

[math] \label{eq:v1} V^{\rm rad}_{\rm out} = w_1 V^{\rm diode\,1}_{\rm out} +w_2 V^{\rm diode\,2}_{\rm out} [/math],

where ω1 and ω2 are inverse-variance weights calculated from the data as discussed in (Planck-Early-V[8]). This way, the diode-diode anti-correlation is cancelled, and the radiometer white noise becomes

[math] \label{eq:dt3} \Delta T^{\rm rad} ≈ \frac{ T_0^{\rm diode}}{\sqrt{2}} (1+\alpha \epsilon_{T{\rm n}})^{1/2}. [/math]

In the equations above, ε«1, while α (a term ≈1) is given by

[math] \label{eq:alpha} \alpha = \frac{ T_{\rm noise} (2 \; T_{\rm noise} + T_{\rm sky} + T_{\rm ref} } {2 (T_{\rm sky} + T_{\rm noise} )(T_{\rm ref} + T_{\rm noise})}. [/math]

See the Diode Combination section for the details of the diode combination procedure.

In Fig. 3 below we show a close-up of the two front end modules of an RCA with the four phase switches, which are labelled with the four letters A and B (main arm), and C and D (side arm). Each phase switch is characterized by two states: state 0 (no phase shift applied to the incoming wave); and state 1 (180° phase shift applied) and can either stay fixed in a state or switch at 4 kHz between the two states.

Phase switches are clocked and biased by the DAE and their configuration can be programmed via telecommand. In order to simplify the instrument electronics, phase switches are configured and operated in pairs, by convention they are labelled A/C (corresponding to the first LNA both of main and side arm) and B/D (corresponding to the second LNA both of main and side arm). This means that if phase switches A and C are switching at 4 kHz then B and D are fixed, both in the same state (either 0 or 1); and viceversa. This simplification, required during the design phase to comply with mass and power budgets, comes at the price of losing some setup redundancy.

Figure 3. Close-up of the two front-end modules of an RCA. There are four phase switches, labelled A, B, C, and D. Each switch can be fixed in one of its two positions (labelled as 0, 1) or switched at 4 kHz between 0 and 1. Phase switches are clocked and biased by the DAE in pairs: A/C and B/D.

Feed Horns (FHs)[edit]

Dual profiled corrugated horns have been selected at all LFI frequencies as the best design in terms of the shape of the main lobe, level of the side lobes, control of the phase centre, and compactness. Details of the design, flight model and tests of Planck-LFI feed horns can be found in (Villa et al. 2009[9]) and in the corresponding Annex section.

Ortho-Mode Transducers (OMTs)[edit]

The Ortho–Mode Transducer (OMTs) separates the radiation collected by the feedhorn into two orthogonal polarization components. It consists of a circular to square waveguide transition (directly connected to the FH), a square waveguide section and two separate rectangular waveguides (the main and side arms, which separate and pick up the orthogonal polarizations, connected with the FEU). A 90° bend is always present on the side arm, while a twist is also necessary on the main (30 and 44 GHz) and side (70 GHz) arms, in order to match the FEU polarization.

The details of the flight models and measurements of the Planck LFI ortho-mode transducers can be found in (D'Arcangelo et al. 2009[10]) and in the corresponding Annex section.

Front End Modules (FEMs)[edit]

Front End Modules are located in the FPU, just behind the Feed Horn and the Ortho Mode Transducers. The 70 GHz FEMs are mounted onto the inner wall of the main frame (the wall facing HFI instrument) from the HFI side. The 44 and 30 GHz FEMs are inserted into the main frame from the waveguide (WG) side and fixed to the bottom plate. Screws to the bottom plate are inserted from the WG side. The LFI FEMs are the first active stage of amplification of the radiometer chain. Each FEM contains four amplification paths, each of which is composed of several cascaded LNAs followed by a phase switch. Two passive hybrids, at the input and output of the FEM, are used to mix pairs of signals of the same radiometer (see Fig. 3). This cuases the instabilities of each chain to be applied to both the sky and load signals.

The passive hybrid coupler ("magic-tee") combines the signals from the sky and cold load with a fixed phase offset of either 90° or 180° between them. It has a 20% bandwidth, low loss, and amplitude balance needed at the output to ensure adequate signal isolation.

The details of the design, development and verification of the 70 GHz front-end modules for the Planck Low Frequency Instrument can be found in (Varis et al. 2009[11]) and in the corresponding Annex section.

Waveguides (WGs)[edit]

The LFI Front End Unit (FEU) is connected to the Back End Unit (BEU) by 44 rectangular waveguides approximately 1.5-2.0 m long. Each waveguide exhibits a low voltage standing wave ratio, low thermal conductivity, low insertion loss, and low mass. In addition, the waveguide path permits the LFI/HFI integration and the electrical bonding between the FPU and the BEU. Because of the Focal Plane Unit arrangement, the waveguides are in general twisted and bent in different planes and with different angles, depending on the particular waveguide. From the thermal point of view the waveguides have to connect two systems (the BEM and FEM) that are at very different temperatures. At BEM level the waveguides are at a temperature of 300 K, while at FEM level the temperature is 20 K. The waveguides have to reduce the thermal flow from 300 K to 20 K. In Fig. 1 in the LFI overview (left panel) a conceptual sketch of the LFI configuration is shown.

Details of the Planck-LFI flight model of the composite waveguides can be found in (D'Arcangelo et al. 2009[1]) and in the corresponding Annex section.

Back End Modules (BEMs)[edit]

The BEMs are composed of four identical channels each made of Low Noise Amplifiers (LNAs), an RF bandpass Filter, RF to DC diode detector, and DC amplifiers. The FEM output signals are connected by waveguides from the Focal Plane Unit (FPU) assembly to the Back End Modules (BEMs) housed adjacent to the Data Acquisition Electronics (DAE) assembly. To maintain compatibility with the FEMs, each BEM accommodates four receiver channels from the four waveguide outputs of each FEM. The BEM internal signal routes are not cross-coupled and can be regarded as four identical parallel circuits. Each BEM is constructed as two mirror halves. The two amplifier/detector assemblies each contain two amplifier/detector circuits. Each is supplied from one of a pair of printed circuit boards, which also house two DC output amplifiers.

The details of the design, development and verification of the 30 and 44 GHz back-end modules for the Planck Low Frequency Instrument can be found in (Artal et al. 2009[12]). Details of the design, development and verification of the 70 GHz back-end modules for the Planck Low Frequency Instrument can be found in (Varis et al. 2009[11]). Details are also reported in the corresponding Annex section.

4K Load[edit]

The purpose of the 4K reference load is to provide the radiometer with a stable reference signal. Reducing the input offset (the radiometric temperature difference between the sky and the reference load) reduces the minimum achievable radiometer 1/f noise knee frequency for a given amplifier fluctuation spectrum. A reference load temperature that matches the sky temperature (approximately 2.7 K) would be ideal.

Details of the design, characteristics, and performance of the LFI 4-K reference load units are given in (Valenziano et al. 2009[2]) and in the corresponding section of the Annexes.

Naming Convention[edit]

The naming of all the LFI elements has been described in the previous sections, but here is summarized again for clarity.

The 11 RCAs are labelled by numbers from 18 to 28, as outlined in Fig. 1 in the LFI overview (right panel). In each RCA, the two perpendicular linear polarization components are labelled as M or S according to the arm of the OMT they are connected to (Main or Side, see lower-left inset of Fig. 2).

Each front-end amplifier (see upper-left inset of Fig. 2) is labelled with the codes 1, 2, so that the four outputs of the FEM LNAs can be named with the sequence M1, M2 (radiometer M) and S1, S2 (radiometer S).

Each radiometer has two output diodes (see upper-right inset of Fig. 2), which are labelled with binary codes 00, 01 (radiometer M) and 10, 11 (radiometer S), so that the four outputs of each radiometric chain can be named with the sequence M-00, M-01, S-10, and S-11.

REBA[edit]

The Radiometer Electronics Box Assembly (REBA) is the electronics unit that processes the digitized scientific data and manages the overall instrument. It is also in charge of communication with the spacecraft. There are two REBA boxes, one nominal and one redundant. The redundancy concept is cold, which means that both boxes are never ON at the same time; the operation of each unit is managed by the spacecraft switching on the corresponding unit. The REBA ASW (Application SoftWare) is the same in each REBA box.

A detailed description of the Planck LFI REBA can be found in (Herreros et al. 2009[3]) and in the corresponding section of the Annexes.

Instrument On-board Software[edit]

The REBA software is the on-board software of LFI. It is installed in the two computing subunits of REBA: the DPU (Digital Processing Unit), responsible of the control and monitoring of the instrument and the interface with the spacecraft; and the SPU (Signal Processing Unit), responsible for the data reduction and compression.

Details can be found in the corresponding section of the Annexes.

Reduction and Compression of Science Data[edit]

To assess stability against 1/f noise, the Low Frequency Instrument (LFI) on board the Planck mission acquired data at a rate much higher than the data rate allowed by the science telemetry bandwidth of 35.5 kbps. The data were processed by an on-board pipeline, followed on the ground by a decoding and reconstruction step, to reduce the volume of data to a level compatible with the bandwidth, while minimizing the loss of information. The on-board processing of the scientific data used by Planck/LFI to fit the allowed data-rate is an intrinsically lossy process which distorts the signal in a manner that depends on a set of five free parameters (Naver, r1, <i,>r2, q, O) for each of the 44 LFI detectors. A brief description of the characteristics of this algorithm and the level of distortion introduced by the on-board processing as a function of these parameters can be found in the corresponding section of the Annexes, while a full description of the Planck LFI on-board data handling system and the tuning and optimization method of the on-board processing chain can be found in (Maris et al. 2009[4]).

The strategy adopted to fit into the telemetry bandwidth relies on three on-board processing steps: downsampling; pre-processing the data to ensure loss-less compression; and loss-less compression itself. To demonstrate these steps, a model of the input signal was used. Note that while the compression is loss-less, the pre-processing is not, due to the need to rescale the data and convert them to integers (a process named data re-quantization). However, the whole strategy is designed to keep strict control over the way in which lossy operations are done, and to quantify the amount of information loss in order to assess the optimal compression rate with minimal information loss.

Instrument Operations[edit]

Operational Modes[edit]

The operations of the LFI are designed to be automatic and require little if any intervention from the ground. A small number of commands are required for operating the instrument and eventually for diagnostic and reconfiguration purposes. Each sky survey is conducted by the LFI with the instrument in the Normal Operations Mode. No deployable elements, or mechanically moving parts are included in the instrument. The scanning of the sky is achieved by progressive repointing of the satellite spin axis, with the Sun direction always within a cone 10° from the spin axis. Within the Normal Science Mode the instrument can be configured in order to fit with different science or diagnostic needs without changing the power consumption and thus the temperature in the FPU.

A brief summary of the LFI Operational Modes and the transitions between them is given in the corresponding section of the Annexes.


Ground Tests[edit]

During its development, the LFI flight model was calibrated and tested at various integration levels from sub-systems (Davis et al. 2009[13]) to individual integrated receivers (Planck-PreLaunch-VI[14]) and the whole receiver array (Planck-PreLaunch-V[15]). In every campaign we performed tests according to the following classification:

  • functionality tests, performed to verify the instrument functionality;
  • tuning tests, to tune radiometer parameters (biases, DC electronics gain and offset, digital quantization and compression) for optimal performance in flight-like thermal conditions;
  • basic calibration and noise performance tests, to characterize instrument performance (photometric calibration, isolation, linearity, noise and stability) in tuned conditions;
  • susceptibility tests, to characterize instrument susceptibility to thermal and electrical variations.

Where possible, the same tests were repeated in several test campaigns, in order to ensure enough redundancy and confidence in the repeatability of the instrument behaviour. A matrix showing the instrument parameters measured in the various test campaigns is provided in Table 1 of (Planck-PreLaunch-V[15]).

The ground test campaign was developed in three main phases: cryogenic tests on the individual RCAs; cryogenic tests on the integrated receiver array (the so-called radiometer array assembly, RAA); and system-level tests after the integration of the LFI and HFI instruments onto the satellite. The first two phases were carried out at the Thales Alenia Space - Italia laboratories located in Vimodrone (Milano, Italy) (note that receiver tests on 70 GHz RCAs were carried out in Finland, at Yilinen laboratories), while system level tests (SLTs) were conducted in a dedicated cryofacility at the Centre Spatiale de Liège (CSL) located in Liege, Belgium.

In Table 1 below we list the temperature of the main cold thermal stages during ground tests compared to in-flight nominal values. These values show that system-level tests were conducted in conditions that were as flight-representative as possible, while results obtained during RCA and RAA tests need to be extrapolated to flight conditions to allow comparison. Details of the RCA test campaign are discussed in (Planck-PreLaunch-VI[14]), while the RAA tests and the extrapolation methods are presented in (Planck-PreLaunch-V[15]).

Table 1. Temperatures of the main cold stages during the various ground test campaigns compared to in-flight nominal values.
Temperature Nominal RCA tests RAA tests System-level
Sky ~ 3 K ≳ 8 K ≳ 18.5 K ~ 4 K
Ref. load ~ 4.5 K ≳ 8 K ≳ 18.5 K ~ 4.5 K
Front-end unit ~ 20 K ~ 20 K ~ 26 K ~ 20 K

During the various test campaigns the instrument was switched off and moved several times in a time period of about three years. A series of functional tests were always repeated at each location and also in-flight, in order to verify the instrument functionality and the response repeatability. No failures or major problems have been identified due to transport and integration procedures.

The expected Planck LFI scientific performance, resulting mainly from cryogenic system level tests, are described in (Planck-PreLaunch-V[15]).

In-flight Calibration[edit]

The LFI Commissioning and Calibration and Performance Verification (CPV) phases started on 4 June 2009 and lasted until 12 August 2009 when Planck started scanning the sky in nominal mode. At the onset of CPV, the active cooling started when the radiating surfaces on the payload module reached their working temperatures (approximately 50 K on the third V-groove, and 40 K on the reflectors) by passive cooling. This was achieved during the transfer phase. Nominal temperatures were achieved on 3 July 2009, when the dilution cooler temperature reached 0.1 K (Planck-Early-II[16], Meinhold et al. 2009 [17]). The cooldown of the HFI 4-K stage (see Fig. 4 below), was key during CPV for the LFI, because it provided a variable input signal that was exploited during bias tuning.

The LFI Commissioning and CPV was carried out in four phases:

  • LFI switch-on and basic functionality verification (Commissioning);
  • tuning of front-end biases and back-end electronics (CPV);
  • preliminary calibration tests (CPV);
  • thermal tests (CPV).
Figure 4. Functional tests performed during CPV against timeline. Curves from thermal sensors monitoring the 4-K stage and the FPU temperature are superimposed.

Details of the LFI Commissioning and CPV test campaign are given in (Gregorio et al. 2013[18]).

Performance Summary[edit]

A summary of the LFI performance parameters is given in Table 1 of the section titled Summary of LFI data characteristics.

Instrument scientific performance[edit]

Optical parameters[edit]

The most accurate measurements of the LFI main beams were made with Jupiter, the most powerful unresolved (to Planck) celestial source in the LFI frequency range. Since the LFI feedhorns point to different positions on the sky, they detect the signal at different times. To map the beam, each sample in the selected timelines was projected in the (u, v) plane perpendicular to the nominal line-of-sight (LOS) of the telescope (and at 85° to the satellite spin axis). The u and v coordinates are defined in terms of the usual spherical coordinates (θ, φ) :

[math] u = \sin θ \, \cos φ, [/math]
[math] v = \sin θ \, \sin φ. [/math]

To increase the signal-to-noise ratio, data were binned in an angular region of 2′ for the 70 GHz channels and 4′ for the 30 and 44 GHz channels. We recovered all beams down to −25 dB from the peak. An elliptical Gaussian was fit to each beam for both M and S radiometers. Differences between the M and S beams caused by optics and receiver non-idealities are inevitable at some level, but they appear to be well within the statistical uncertainties, and for the purposes of point-source extraction, the beams may be considered identical. For the details of the typical FWHM and ellipticity averaged over each frequency channel, refer to (Planck-Early-III[19]), (Planck-2015-A05[20]) and Planck-2020-A2[21]. Exhaustive details on all LFI beam parameters are presented in the LFI data processing section entitled Beams.

Photometric Calibration[edit]

Photometric calibration, i.e., conversion from voltage to antenna temperature, is performed for each radiometer after total power data have been cleaned of 1 Hz frequency spikes (see the LFI data processing section Spikes Removal page and Planck-Early-V[8], Planck-2015-A03[22], Planck-2020-A2[21] and differenced.

Here we report a brief overview of the photometric calibration; for details see the LFI data processing section Photometric Calibration.

Our calibrator is the dipole signal induced by spacecraft motion with respect to the CMB rest frame. The largest calibration uncertainty comes from the presence of the Galaxy and of the CMB anisotropies in the measured signal. We therefore use an iterative calibration procedure in which the dipole is fitted and subtracted, producing a sky map that is then removed from the original data to enhance the dipole signal for the next iteration. Typically, convergence is obtained after a few tens of iterations.

We calibrate in two major steps. First, we find the absolute calibration using the motion of the spacecraft around the sun (the 'orbital dipole'). This process uses only the temperature of the CMB and the very well known velocity of the spacecraft as inputs. From this calibration, and comparing with HFI, we can find the Planck estimate of the solar system velocity relative to the CMB (the 'solar dipole'). Second, We run another iterative calibration with this solar dipole as our signal. The larger signal gives us a better estimate of the time variation of our gain than that obtained from the orbital dipole.

Noise Properties[edit]

The noise characteristics of the LFI data streams are closely reproduced by a simple (white + 1/f ) noise model,

[math] \label{eq:p1} P(f)= \sigma^2 \left[1+ \left(\frac{f}{f_{\rm knee}}\right)^\alpha \right] [/math],

where P(f) is the power spectrum and α≈−1. In this model, noise properties are characterized by three parameters, the white noise limit σ, the knee frequency fknee, and the exponent of the 1/f component α, also referred to as the slope. Here we give noise performance estimates based on one year of operations; details of the analysis are given in LFI data processing section (Noise estimation) and in (Planck-Early-V[8], Planck-Early-III[19], Planck-2015-A03[22] and Planck-2020-A2[21].

Noise properties have been calculated following two different and complementary approaches: 1) fitting the equation above to time-ordered data for each radiometer; and 2) building normalized noise maps by differencing data from the first half of each pointing period with data from the second half of that pointing period to remove the sky signal (“jackknife” data sets).

Typical uncertainties are 0.5% for the white noise, between 5 and 10% for the slope, and between 10 and 20% for the knee frequency.

White Noise Sensitivity[edit]

Details of the white noise sensitivity can be found in (Planck-Early-III[19]).

Table 2 summarizes the sensitivity numbers calculated during the first year of operations using methods and procedures described in detail in (Planck-Early-V[8], Planck-2015-A03[22] and Planck-2020-A2[21], compared with scientific requirements. The measured sensitivity is in very good agreement with pre-launch expectations. While the white noise moderately exceeds the design specification, this performance is fully in line with the LFI science objectives.

Table 2. White noise sensitivity of the LFI frequency channels compared with requirements.
Channel Measurement [μKCMB s1/2] Requirement [μKCMB s1/2]
70 GHz 152.0 119
44 GHz 174.2 119
30 GHz 148.1 119

Instrument technical performance[edit]

Spectral response[edit]

The in-band receiver response was thoroughly modelled and measured for all the LFI detectors during ground tests. The complete set of bandpass curves was published in (Zonca et al. 2009[23], Planck-2015-A06[24]) where all the details of the LFI radiometer's spectral response are given. From each curve we have derived the effective centre frequency according to:

[math] \label{eq:spectr} \nu_0 = \frac{ \int_{\nu_{\rm min}}^{\nu_{\rm max}} \nu g(\nu) \; d \nu} {\int_{\nu_{\rm min}}^{\nu_{\rm max}} g(\nu)\; d\nu } [/math]

where Δν = νmax − νmin is the receiver bandwidth and g(ν) is the bandpass response. Details about colour corrections, C(α), needed to derive the brightness temperature of a source with a power-law spectral index α, are provided in

Some details are also given in the corresponding section of the Annexes.


Bandpass estimation[edit]

As detailed in (Zonca et al. 2009[23], Planck-2015-A06[24]), our most accurate method to measure the LFI bandpasses is based on measurements of individual components integrated into the LFI Advanced RF Model (LARFM) to yield a synthesized radiometer bandpass. The LARFM is a software tool based on the open-source Quasi Universal Circuit Simulator (QUCS). The measured frequency responses of the various subsystems (feed-OMT, FEM, BEM) are considered as lumped S-parameter components. Measurements of single components are obtained with standard methods and provide highly reliable results, with precision of order 0.1-0.2 dB over the entire band. Waveguides are simulated with an analytical model, in order to reproduce the effect of their temperature gradient and the effect of standing waves caused by impedance mismatch at the interfaces between the FEM and BEM. This is because the 1.8-m long waveguides were not measured at unit level in cryogenic conditions. The model provides accurate agreement with the measured waveguide response in the conditions of the test measurements (300 K). The composite bandpasses are estimated to have a precision of about 1.5 to 2 dB.

Some details are also given in the corresponding section of the Annexes.

Stability[edit]

Thanks to its differential scheme, the LFI is insensitive to many effects caused by 1/f noise, thermal fluctuations, or electrical instabilities. As detailed in (Planck-Early-III[19]), one effect detected during the first survey was the daily temperature fluctuation in the back-end unit induced by the downlink transponder, which was powered on each day for downlinks during the first 258 days of the mission. As expected, the effect is highly correlated between the sky and reference load signals. In the difference, the variation is reduced by a factor around (1 − r), where r is the gain modulation factor defined above (see r definition in RCA section).

A particular class of signal fluctuations occasionally observed during operations was due to electrical instabilities that appeared as abrupt increases in the measured drain current of the front-end amplifiers, with a relaxation time variable from few seconds to some hundreds of seconds. Typically, these events caused a simultaneous change in the sky and reference load signals. Because they are essentially common-mode, their residual on the differenced data is negligible, and the data are suitable for science production. In a few cases the residual fluctuation in the differential output was large enough (a few millikelvin in calibrated antenna temperature units) to be flagged, and the data were not used. The total amount of discarded data for all LFI channels until Operational Day 389 was about 2000s per detector, or 0.008%.

A further peculiar effect appeared in the 44 GHz detectors, where single isolated samples, either on the sky or the reference voltage output, were clear outliers compared with the rest. Over a reference period of four months, 15 occurrences of single-sample spikes (out of 24 total anomaly events) were discarded, an insignificant loss of data.

Thermal susceptibility[edit]

As mentioned in the section LFI In-flight Calibration above, and detailed in (Gregorio et al. 2013[18]), during the CPV campaign, susceptibility tests were performed in order to characterize the LFI instrument susceptibility to thermal and electrical variations.

The effect of temperature fluctuations on the LFI radiometers originates in the Planck cold end interface of the hydrogen sorption cooler to the instrument focal plane. The temperature is actively controlled through a dedicated stage, the Thermal Stabilization Assembly (TSA), providing a first reduction of the effect. The thermal mass of the focal plane strongly contributes to reduce residual fluctuations. The physical temperature fluctuations propagated at the front end modules cause a correlated fluctuation in the radiometer signal, degrading the quality of scientific data. The accurate characterization of this effect is crucial for attempts to remove it from raw data by exploiting the housekeeping information on thermal sensors.

The propagation of the temperature oscillations through the focal plane and the instrument response to thermal changes were characterized through two main tests:

  • the thermal dynamic response, aimed at measuring the dynamic thermal behaviour of the LFI Focal Plane;
  • the thermal susceptibility of the radiometers.

Further details are also given in the corresponding section of the Annexes.

Instrument budgets[edit]

LFI power, mass and telemetry budgets are given in the corresponding section of the Annexes.

Systematic Effects [edit]

The LFI design was driven by the need to suppress systematic effects well below instrument white noise. The differential receiver scheme, with reference loads cooled to 4 K, greatly minimizes the effect of 1/f noise and common-mode fluctuations, such as thermal perturbations in the 20-K LFI focal plane. The use of a gain modulation factor (see r definition in the RCA section above) largely compensates for spurious contributions from input offsets. Furthermore, diode averaging (see Vradout definition in the RCA section above) allows us to cancel second-order correlations, such as those originating from phase switch imbalances.

We have developed an error budget for systematic effects (Planck-PreLaunch-IV[7], Planck-Early-III[19]Planck-2013-III[25]Planck-2015-A04[26] and Planck-2020-A2[21]) as a reference for both instrument design and data analysis. Our goal is to ensure that each systematic effect is rejected to the specified level, either by design or by robust removal in software. At this stage, the following effects are relevant:

Selection 258.png

For each of these effects we used flight data and information from ground tests to build timelines, maps, and angular power spectra that represent our best knowledge of their impact on the scientific analysis. The details of the systematic effect analysis are given in the section entitled LFI data processing section (Systematic Effects uncertainties).

The Sorption Cooler[edit]

The Planck H2 Sorption Cooler is the first stage of the active cryogenic chain. Its task is to maintain the LFI down to the operating temperature while providing a pre-cooling stage for the HFI refrigerators. The system performs a simple thermodynamic cycle based on hydrogen compression, gas pre-cooling by three passive radiators, further cooling due to the heat recovery by the cold low pressure gas stream, expansion through a J-T valve, and evaporation at the cold stage (Planck-Early-II[16]). A schematic of the Planck Sorption Cooler System (SCS) is shown in Fig. 5. The engine of the cryocooler is the compressor. It serves two main functions: to produce the high-pressure hydrogen gas flow; and to maintain a stable gas recovery rate, which keeps the return pressure, hence the liquid temperature, constant. The high pressure gas flowing to the cold end is pre-cooled by exchanging heat with the three passive stages at the V-Grooves and with the evaporated cold gas returning back to the compressor. The gas then expands at the cold end through a J-T valve, producing approximately 1 Watt of cooling power at a temperature of <20K: most of this heat lift is used at the LVHX2 (Liquid Vapour Heat eXchanger 2, see Fig. 5) interface to absorb the LFI heat load at a temperature around 20K. The remaining heat lift is used at the LVHX1 as a pre-cooling stage, at a temperature lower than 19 K, for the two HFI refrigerators. The cooler and its performance are described in detail in (Morgante et al. 2009[27]).

Figure 5. Planck Sorption Cooler CAD view with labels of the main sub-systems.

Both LVHXs provide a temperature of around 18 K, with fluctuations driven by the cooler instabilities (compressor element variations, cycling, two-phase flow dynamics, etc.). Stabilization of the HFI interface (LVHX1) temperature is not necessary, since thermal control of the subsequent colder stages is more efficient and very effective. To reduce cold end fluctuations directly transmitted to the LFI radiometers, a copper block, designated as the Temperature Stabilization Assembly (TSA), is inserted, as an intermediate stage, between LVHX2 and the LFI Focal Plane Unit (FPU). The TSA comprises a temperature sensor and a heater, controlled by a PID-type feedback loop, working in combination with the passive thermal inertia. The set-point temperature of the TSA is an adjustable parameter of the sorption cooler system, chosen to provide dynamic range for control during mission: as the compressor elements age, the return gas pressure and thus the temperature of LVHX2 rise slowly. To keep the LFI temperature reference stable during operations the set point must be periodically adjusted to maintain the level of oscillations within the required range (Fig. 6).

Figure 6. LVHX2 (black, HK parameter name SD029540) and TSA (red, HK parameter name SD030540) temperature profile from 125 to 563 days after launch.

Two sorption cooler units were integrated on-board the Planck spacecraft. Both units were used to cover the mission lifetime, with one cooler operated for the first two sky surveys. At the end of its lifetime a switchover operation, performed on August 11th 2010 (455 days after launch), activated the second unit that ran since then. The LVHX2 temperature change due to the new cooler start, and the subsequent re-adjustment of the LFI temperature stabilization stage, is clear in Fig. 6.


References[edit]

  1. Jump up to: 1.01.1 The Planck-LFI flight model composite waveguides, O. D'Arcangelo, L. Figini, A. Simonetto, F. Villa, M. Pecora, P. Battaglia, M. Bersanelli, R. C. Butler, F. Cuttaia, S. Garavaglia, P. Guzzi, N. Mandolesi, A. Mennella, G. Morgante, L. Pagan, L. Valenziano, Journal of Instrumentation, 4, 2007-+, (2009).
  2. Jump up to: 2.02.1 Planck-LFI: design and performance of the 4 Kelvin Reference Load Unit, L. Valenziano, F. Cuttaia, A. De Rosa, L. Terenzi, A. Brighenti, G. P. Cazzola, A. Garbesi, S. Mariotti, G. Orsi, L. Pagan, F. Cavaliere, M. Biggi, R. Lapini, E. Panagin, P. Battaglia, R. C. Butler, M. Bersanelli, O. D'Arcangelo, S. Levin, N. Mandolesi, A. Mennella, G. Morgante, G. Morigi, M. Sandri, A. Simonetto, M. Tomasi, F. Villa, M. Frailis, S. Galeotta, A. Gregorio, R. Leonardi, S. R. Lowe, M. Maris, P. Meinhold, L. Mendes, L. Stringhetti, A. Zonca, A. Zacchei, Journal of Instrumentation, 4, 2006-+, (2009).
  3. Jump up to: 3.03.1 The Planck-LFI Radiometer Electronics Box Assembly, J. M. Herreros, M. F. Gómez, R. Rebolo, H. Chulani, J. A. Rubiño-Martin, S. R. Hildebrandt, M. Bersanelli, R. C. Butler, M. Miccolis, A. Peña, M. Pereira, F. Torrero, C. Franceschet, M. López, C. Alcalá, Journal of Instrumentation, 4, 2008-+, (2009).
  4. Jump up to: 4.04.1 Optimization of Planck-LFI on-board data handling, M. Maris, M. Tomasi, S. Galeotta, M. Miccolis, S. Hildebrandt, M. Frailis, R. Rohlfs, N. Morisset, A. Zacchei, M. Bersanelli, P. Binko, C. Burigana, R. C. Butler, F. Cuttaia, H. Chulani, O. D'Arcangelo, S. Fogliani, E. Franceschi, F. Gasparo, F. Gomez, A. Gregorio, J. M. Herreros, R. Leonardi, P. Leutenegger, G. Maggio, D. Maino, M. Malaspina, N. Mandolesi, P. Manzato, M. Meharga, P. Meinhold, A. Mennella, F. Pasian, F. Perrotta, R. Rebolo, M. Türler, A. Zonca, Journal of Instrumentation, 4, 2018-+, (2009).
  5. Jump up to: 5.05.1 1/f noise and other systematic effects in the Planck-LFI radiometers, M. Seiffert, A. Mennella, C. Burigana, N. Mandolesi, M. Bersanelli, P. Meinhold, P. Lubin, A&A, 391, 1185-1197, (2002).
  6. Jump up to: 6.06.1 Offset balancing in pseudo-correlation radiometers for CMB measurements, A. Mennella, M. Bersanelli, M. Seiffert, D. Kettle, N. Roddis, A. Wilkinson, P. Meinhold, A&A, 410, 1089-1100, (2003).
  7. Jump up to: 7.07.1 Planck pre-launch status: Design and description of the Low Frequency Instrument, M. Bersanelli, N. Mandolesi, R. C. Butler, et al. , A&A, 520, A4+, (2010).
  8. Jump up to: 8.08.18.28.38.4 Planck early results. V. The Low Frequency Instrument data processing, A. Zacchei, D. Maino, C. Baccigalupi, et al. , A&A, 536, A5, (2011).
  9. Jump up Planck-LFI flight model feed horns, F. Villa, O. D'Arcangelo, M. Pecora, L. Figini, R. Nesti, A. Simonetto, C. Sozzi, M. Sandri, P. Battaglia, P. Guzzi, M. Bersanelli, R. C. Butler, N. Mandolesi, Journal of Instrumentation, 4, 2004-+, (2009).
  10. Jump up The Planck-LFI flight model ortho-mode transducers, O. D'Arcangelo, A. Simonetto, L. Figini, E. Pagana, F. Villa, M. Pecora, P. Battaglia, M. Bersanelli, R. C. Butler, S. Garavaglia, P. Guzzi, N. Mandolesi, C. Sozzi, Journal of Instrumentation, 4, 2005-+, (2009).
  11. Jump up to: 11.011.1 Design, development, and verification of the Planck Low Frequency Instrument 70 GHz Front-End and Back-End Modules, J. Varis, N. J. Hughes, M. Laaninen, V.-H. Kilpiä, P. Jukkala, J. Tuovinen, S. Ovaska, P. Sjöman, P. Kangaslahti, T. Gaier, R. Hoyland, P. Meinhold, A. Mennella, M. Bersanelli, R. C. Butler, F. Cuttaia, E. Franceschi, R. Leonardi, P. Leutenegger, M. Malaspina, N. Mandolesi, M. Miccolis, T. Poutanen, H. Kurki-Suonio, M. Sandri, L. Stringhetti, L. Terenzi, M. Tomasi, L. Valenziano, Journal of Instrumentation, 4, 2001-+, (2009).
  12. Jump up LFI 30 and 44 GHz receivers Back-End Modules, E. Artal, B. Aja, M. L. de la Fuente, J. P. Pascual, A. Mediavilla, E. Martinez-Gonzalez, L. Pradell, P. de Paco, M. Bara, E. Blanco, E. García, R. Davis, D. Kettle, N. Roddis, A. Wilkinson, M. Bersanelli, A. Mennella, M. Tomasi, R. C. Butler, F. Cuttaia, N. Mandolesi, L. Stringhetti, Journal of Instrumentation, 4, 2003-+, (2009).
  13. Jump up Design, development and verification of the 30 and 44 GHz front-end modules for the Planck Low Frequency Instrument, R. J. Davis, A. Wilkinson, R. D. Davies, W. F. Winder, N. Roddis, E. J. Blackhurst, D. Lawson, S. R. Lowe, C. Baines, M. Butlin, A. Galtress, D. Shepherd, B. Aja, E. Artal, M. Bersanelli, R. C. Butler, C. Castelli, F. Cuttaia, O. D'Arcangelo, T. Gaier, R. Hoyland, D. Kettle, R. Leonardi, N. Mandolesi, A. Mennella, P. Meinhold, M. Pospieszalski, L. Stringhetti, M. Tomasi, L. Valenziano, A. Zonca, Journal of Instrumentation, 4, 2002-+, (2009).
  14. Jump up to: 14.014.1 Planck pre-launch status: Calibration of the Low Frequency Instrument flight model radiometers, F. Villa, L. Terenzi, M. Sandri, et al., A&A, 520, A6+, (2010).
  15. Jump up to: 15.015.115.215.3 Planck pre-launch status: Low Frequency Instrument calibration and expected scientific performance, A. Mennella, M. Bersanelli, R. C. Butler, et al. , A&A, 520, A5+, (2010).
  16. Jump up to: 16.016.1 Planck early results. II. The thermal performance of Planck, Planck Collaboration II, A&A, 536, A2, (2011).
  17. Jump up Noise properties of the Planck-LFI receivers, P. Meinhold, R. Leonardi, B. Aja, E. Artal, P. Battaglia, M. Bersanelli, E. Blackhurst, C. R. Butler, L. P. Cuevas, F. Cuttaia, O. D'Arcangelo, R. Davis, M. L. de la Fuente, M. Frailis, C. Franceschet, E. Franceschi, T. Gaier, S. Galeotta, A. Gregorio, R. Hoyland, N. Hughes, P. Jukkala, D. Kettle, M. Laaninen, P. Leutenegger, S. R. Lowe, M. Malaspina, R. Mandolesi, M. Maris, E. Martínez-González, L. Mendes, A. Mennella, M. Miccolis, G. Morgante, N. Roddis, M. Sandri, M. Seiffert, M. Salmón, L. Stringhetti, T. Poutanen, L. Terenzi, M. Tomasi, J. Tuovinen, J. Varis, L. Valenziano, F. Villa, A. Wilkinson, F. Winder, A. Zacchei, A. Zonca, Journal of Instrumentation, 4, 2009-+, (2009).
  18. Jump up to: 18.018.1 In-flight calibration and verification of the Planck-LFI instrument, A. Gregorio, F. Cuttaia, A. Mennella, M. Bersanelli, S. Maris, P. Meinhold, Submitted to Journal of Instrumentation, (2013).
  19. Jump up to: 19.019.119.219.319.4 Planck early results. III. First assessment of the Low Frequency Instrument in-flight performance, A. Mennella, R. C. Butler, A. Curto, et al. , A&A, 536, A3, (2011).
  20. Jump up Planck 2015 results. IV. LFI beams and window functions, Planck Collaboration, 2016, A&A, 594, A4.
  21. Jump up to: 21.021.121.221.321.4 Planck 2018 results. II. Low Frequency Instrument data processing, Planck Collaboration, 2020, A&A, 641, A2.
  22. Jump up to: 22.022.122.2 Planck 2015 results. II. LFI processing, Planck Collaboration, 2016, A&A, 594, A2.
  23. Jump up to: 23.023.1 Planck-LFI radiometers' spectral response, A. Zonca, C. Franceschet, P. Battaglia, F. Villa, A. Mennella, O. D'Arcangelo, R. Silvestri, M. Bersanelli, E. Artal, R. C. Butler, F. Cuttaia, R. J. Davis, S. Galeotta, N. Hughes, P. Jukkala, V.-H. Kilpiä, M. Laaninen, N. Mandolesi, M. Maris, L. Mendes, M. Sandri, L. Terenzi, J. Tuovinen, J. Varis, A. Wilkinson, Journal of Instrumentation, 4, 2010-+, (2009).
  24. Jump up to: 24.024.1 Planck 2015 results. V. LFI calibration, Planck Collaboration, 2016, A&A, 594, A5.
  25. Jump up Planck 2013 results. III. Low Frequency Instrument systematic uncertainties, Planck Collaboration, 2014, A&A, 571, A3.
  26. Jump up Planck 2015 results. III. LFI systematics, Planck Collaboration, 2016, A&A, 594, A3.
  27. Jump up Cryogenic characterization of the Planck sorption cooler system flight model, G. Morgante, D. Pearson, F. Melot, P. Stassi, L. Terenzi, P. Wilson, B. Hernandez, L. Wade, A. Gregorio, M. Bersanelli, C. Butler, N. Mandolesi, Journal of Instrumentation, 4, 2016-+, (2009).

LFI warm electronics Back End Unit

(Planck) Low Frequency Instrument

High Electron Mobility Transistor

(Planck) High Frequency Instrument

LFI Radiometer Array Assembly

Sorption Cooler Subsystem (Planck)

LFI Radiometer Electronics Box Assembly

LFI cryogenic amplifying stage Front End Unit

Service Module

Sorption Cooler Compressor assembly

Focal Plane Unit

LFI warm electronics Back End Module

LFI Data Acquisition Electronics

revolutions per minute

LFI Radiometer Chain Assembly

LFI cryogenic amplifying stage Front End Module

LFI Ortho Module Transducer

analog to digital converter

Feed Horn

LFI Waveguide

Data Processing Unit

Signal Processing Unit

Centre Spatial de Liège

Calibration and Performance Verification

Line Of Sight

Full-Width-at-Half-Maximum

Cosmic Microwave background

hermal Stabilization Assembly

House Keeping