Difference between revisions of "Summary LFI"

From Planck Legacy Archive Wiki
Jump to: navigation, search
 
(15 intermediate revisions by 4 users not shown)
Line 1: Line 1:
 
{{DISPLAYTITLE:Summary of LFI data characteristics}}
 
{{DISPLAYTITLE:Summary of LFI data characteristics}}
The LFI performance are summarized in Table 1 below. All the details can be found in the LFI data processing paper {{BibCite|planck2013-p02}} {{P2013|2}}.
+
Quantification of the performance of LFI is summarized in Table 1 below. All the details can be found in the LFI data processing paper {{PlanckPapers|planck2016-l02}}.  
  
 +
{| border="1" cellspacing="0" cellpadding="2" align="center"
 +
|+ '''Table 1. LFI performance parameters'''
 +
|-
 +
!scope="col"| Parameter
 +
!scope="col"| 30 GHz
 +
!scope="col"| 44 GHz
 +
!scope="col"| 70 GHz
 +
|-
 +
|width="320" | Centre frequency [GHz]
 +
|width="100" | 28.4
 +
|width="100" | 44.1
 +
|width="100" | 70.4
 +
|-
 +
|width="320" | Bandwidth [GHz]
 +
|width="100" | 9.89
 +
|width="100" | 10.72
 +
|width="100" | 14.90
 +
|-
 +
|width="320" | Scanning beam FWHM<sup>a</sup> [arcmin]
 +
|width="100" | 33.10
 +
|width="100" | 27.94
 +
|width="100" | 13.08
 +
|-
 +
|width="320" | Scanning beam ellipticity<sup>a</sup>
 +
|width="100" | 1.37
 +
|width="100" | 1.25
 +
|width="100" | 1.27
 +
|-
 +
|width="320" | Effective beam FWHM<sup>b</sup> [arcmin]
 +
|width="100" | 32.29
 +
|width="100" | 27.00
 +
|width="100" | 13.22
 +
|-
 +
|width="320" | White noise level in timelines<sup>c</sup> [&mu;K<sub>CMB</sub> s<sup>1/2</sup>]
 +
|width="100" | 147.9
 +
|width="100" | 174.0
 +
|width="100" | 151.9
 +
|-
 +
|width="320" | <i>f</i><sub>knee</sub><sup>c</sup> [mHz]
 +
|width="100" | 113.9
 +
|width="100" | 53.0
 +
|width="100" | 19.6
 +
|-
 +
|width="320" | 1/<i>f</i> slope<sup>c</sup>
 +
|width="100" | -0.92
 +
|width="100" | -0.88
 +
|width="100" | -1.20
 +
|-
 +
|width="320" | Overall calibration uncertainty<sup>d</sup> [%]
 +
|width="100" | 0.17
 +
|width="100" | 0.12
 +
|width="100" | 0.20
 +
|-
 +
|}
 +
 +
 
 +
 +
<sup>a</sup> Determined by fitting Jupiter observations directly in the timelines.
 +
 +
<sup>b</sup> Calculated from the main beam solid angle of the effective beam, &Omega;<sub>eff</sub> = mean(&Omega;). These values are used in the source extraction pipeline ({{PlanckPapers|planck2014-a35}}).
 +
<!-- {{PlanckPapers|planck2013-p05}} -->
 +
 +
<sup>c</sup> Typical values derived from fitting noise spectra
 +
 +
<sup>d</sup> difference between first and last iteration of the iterative calibration (for 30 and 44 GHz) or E2E 2015 result (for 70 GHz). In 2015 the calibration uncertainity was 0.35% and 0.26% at 30 and 44, respectively.
  
<center>
 
'''Table 1. Summary of the LFI performance parameters'''
 
[[File:tab_perf2.jpg|thumb|center|600px|
 
<math>^{ \rm a \, }</math> FWHM and ellipticity from scanning beam evaluated fitting Jupiter directly in the timelines. <math>\hspace{480mm} </math>
 
<math>^{ \rm b \, }</math> FWHM from effective beam estimated from the main beam solid angle of the effective beam. Those are the values used in the source extraction pipeline <cite>#planck2013-p05</cite> {{P2013|28}}.
 
<math>{\hspace{35mm}}</math>
 
<math>^{ \rm c \, }</math> White noise per pixel computed from half-ring difference maps. Those values are within the 1% with the white noise sensitivity computed directly on the timelines, taking in account the actual mission time minus the manoeuvres and bad-science flagged data.
 
]]
 
</center>
 
  
 
<!--
 
<!--
Line 25: Line 81:
  
 
== References ==
 
== References ==
 
+
<references />
<biblio force=false>
 
#[[References]]
 
</biblio>
 
  
 
[[Category:LFI data processing|008]]
 
[[Category:LFI data processing|008]]

Latest revision as of 16:00, 5 July 2018

Quantification of the performance of LFI is summarized in Table 1 below. All the details can be found in the LFI data processing paper Planck-2020-A2[1].

Table 1. LFI performance parameters
Parameter 30 GHz 44 GHz 70 GHz
Centre frequency [GHz] 28.4 44.1 70.4
Bandwidth [GHz] 9.89 10.72 14.90
Scanning beam FWHMa [arcmin] 33.10 27.94 13.08
Scanning beam ellipticitya 1.37 1.25 1.27
Effective beam FWHMb [arcmin] 32.29 27.00 13.22
White noise level in timelinesc [μKCMB s1/2] 147.9 174.0 151.9
fkneec [mHz] 113.9 53.0 19.6
1/f slopec -0.92 -0.88 -1.20
Overall calibration uncertaintyd [%] 0.17 0.12 0.20


a Determined by fitting Jupiter observations directly in the timelines.

b Calculated from the main beam solid angle of the effective beam, Ωeff = mean(Ω). These values are used in the source extraction pipeline (Planck-2015-A26[2]).

c Typical values derived from fitting noise spectra

d difference between first and last iteration of the iterative calibration (for 30 and 44 GHz) or E2E 2015 result (for 70 GHz). In 2015 the calibration uncertainity was 0.35% and 0.26% at 30 and 44, respectively.


References[edit]

(Planck) Low Frequency Instrument

Full-Width-at-Half-Maximum

Cosmic Microwave background